Electrical measuring instruments >> calibrators, compartors, calibration rigs

N4-101 (H4-101)

389 704 RUB
101 923 UAH

Online prices include VAT.

For order is possible effectively delivery of any quantity.

Universal calibrator N4-101

Also, this device can be called: N4101, N4 101, h4-101, h4101, h4 101.

N4-101 universal calibrator - used to play the voltage and the DC and AC currents.

The basis of operation of the device N4-101 converted numerical value is reproduced signal into an analog signal by digital-to-analogue converter and subsequent formation of the signal output voltage or current.

Scope: verification and control of electrical appliances direct and alternating currents.

Specifications N4-101:

DC voltage:

Range - 50 mkV to 1000 V.

Subbands - 200 mV, 2 V, 20 V, 200 V, 1000 V.

Tolerance - from 0,02% to 0,05%.

Voltage ripple and noise - from 1 mV to 250 mV.

Maximum load current - 5 mA to 10 mA.

AC voltage of sinusoidal form:

Range - from 1 mV to 750 V.

Subbands - 200 mV, 2 V, 20 V, 200 V, 750 V.

Tolerance - from 0,15% to 0,2%.

Maximum load current - 5 mA to 10 mA.

Load Capacity - 1 pF to 250 pF.

THD - from 0,25% to 0,5%.

Current DC:

Range N4-101 - 0,05 mA to 50 A.

Subbands - 200 mA, 2 mA, 20 mA, 200 mA, 2 A, 50 A.

Error calibrator - from 0,03% to 0,15%.

Maximum load voltage - 1 V to 5 V.

Noise and ripple voltage at the maximum voltage across the load in a range from 2 A to 50 A - no more than 10 mV.

Power AC pure sine wave:

Range - 10 mA to 50 A.

Subbands - 2 mA to 20 mA, 200 mA, 2 A, 50 A.

Tolerance - from 0,15% to 0,2%.

Maximum load voltage - 0,7 V to 3 V.

THD N4-101 - from 0,25% to 1%.

Readability (frequency range - from 40 Hz to 2500 Hz) - 1 Hz.

Error of frequency - no more than ±1%.

Production and operating margin - 20% of the basic error.

Calibrator Interface - RS-232.

Speed ​​of information exchange - 19200 bits/s 57600 bits/s.

Continuous operation - no less than 24 hours.

Dimensions - 540,5x465x184,5 mm.

Weight - 20 kg.

Operating conditions N4-101:

Ambient temperature

- Operating - from 5º C to 40º C;

- Norm - 20±5° C.

Relative humidity at 25º C:

- Working - up to 80%;

- Rate - from 30% to 80%.

Atmospheric pressure - from 106,7 kPa to 84 kPa (630 mm hg to 800 mm hg).

AC voltage - 230±23 V.

The frequency of the AC - 50 Hz±0,5.

Power consumption devices calibrators - 550 V·A.

Antidecoying - Class V.

MTBF - not less than 15000 hours.

Average yield N4-101 - at least 15000.

MTTR - less than 4 hours.

Photos: N4-101 (H4-101)

N4-101 device image.
N4-101 device image.


Яндекс цитирования