Laboratory equipment >> metal hardness testers

UT 5011A Universal device for measuring the hardness method Rockwell, Brinell, Vickers VT 5011A.

UT 5011A

For order is possible effectively delivery of any quantity.
Universal device for measuring the hardness method Rockwell, Brinell, Vickers UT 5011A (UT5011A, UT 5011 A, UT-5011-A, UT5011-A, UT5011 A, UT-5011A)
 
Universal device for measuring the hardness method Rockwell, Brinell, Vickers UT 5011A designed to measure the hardness of metals and alloys?plastics, graphite and metallografitov, plywood, particle board and other materials by the method of Rockwell, Brinell, Vickers.
The device is designed for use in workshops and laboratories of engineering and metallurgical enterprises, as well as in laboratories of research institutes with an ambient temperature of +10 º C to +35 º C, relative humidity from 50% to 80%.
 
Technical characteristics of the instrument versatile instruments for measuring the hardness of the method of Rockwell, Brinell, Vickers:
 
Range of hardness measurements - 20HRA-88HRA, 20HRB-100HRB, 20HRC-70HRC, 32HBW-200HBW? 95HBW-650HBW;
The test load device universal instrument for measuring the hardness of the method of Rockwell, Brinell, Vickers UT 5011A, pre - 98.07 N;
Test loads, general - 588.4 N, 980.7 N, 1471N, 612.9 N, 1839N, 294.2 N, 980.7 N;
The maximum permissible errors of test loads device universal instrument for measuring the hardness of the method of Rockwell, Brinell, Vickers UT 5011A - ± 0,5%, ± 1,0%;
Measurement range of Rockwell hardness according to the values ​​listed in Table 1;
 
Table 1: Range of hardness instrument versatile instrument for measuring the hardness method Rockwell, Brinell, Vickers UT 5011A HRC
Hardness scale
Load, N
View of the indenter
Measuring range, HR
NRA
588.4 N
Diamond cone
from 20 to 88
NRB
980.7 N
Ball Ø1, 588mm
20 to 100
HRC
1471N
Diamond cone
20 to 70
 
Range of hardness measurement device universal instrument for measuring the hardness of the method of Rockwell, Brinell, Vickers UT 5011A by Brinell in line with the values ​​listed in Table 2;
 
Table 2 Range of hardness measurement device universal instrument for measuring the hardness of the method of Rockwell, Brinell, Vickers UT 5011A
Hardness scale
Load, N
View of the indenter
Measuring range, NBW
HBW 2,5 / 62,5
612.9 N
Ball Ø2, 5mm
32-200
HBW 2,5 / 187,5
1839N
Ball Ø2, 5mm
95-650
 
Measurement range of hardness on the Vickers scale 50 - 1500HV;
Test load device universal instrument for measuring the hardness of the method of Rockwell, Brinell, Vickers UT 5011A:
- Rockwell method:
- Preview - 98.07 N;
- Full - 588.4 N; 980.7 N; 1471N;
- Brinell method - 612.9 N; 1839N;
- By Vickers - 294.2 N; 980.7 N;
The maximum permissible errors of test loads device universal instrument for measuring the hardness of the method of Rockwell, Brinell, Vickers UT 5011A:
- Rockwell method:
- Pre - ± 1%;
- Full - ± 0,5%;
- Brinell method - ± 1%;
- By Vickers - ± 0,5%;
Limits of permissible absolute error of measurement of hardness instrument versatile instrument for measuring the hardness method Rockwell, Brinell, Vickers UT 5011A Rockwell are shown in Table 3.
 
Table 3: Limits of the absolute error of hardness measurement device universal instrument for measuring the hardness method Rockwell, Brinell, Vickers UT 5011A HRC
Intervals Hardness Rockwell c designation
Limits of permissible absolute error of measurement of hardness HRC, HR, (±)
from 20HRA to 75HRA
± 2.0HR
from 75HRA to 88HRA
± 1.5HR
from 20HRB to 80HRB
± 3.0HR
from 80HRB to 100HRB
± 2.0HR
from 20HRC to 35HRC
± 2.0HR
from 35HRC to 55HRC
± 1.5HR
from 55HRC to 70HRC
± 1.0HR
 
Limits of permissible absolute error of measurement of hardness instrument versatile instrument for measuring the hardness method Rockwell, Brinell, Vickers UT 5011A on the Vickers scale are shown in Table 4;
 
Table 4: Maximum permissible absolute error of hardness measurement device universal instrument for measuring the hardness of the method of Rockwell, Brinell, Vickers UT 5011A on the Vickers scale
Marking scales Hardness
Measurement intervals Vickers Hardness, HV
100 ± 50
200 ± 50
300 ± 50
400 ± 50
500 ± 50
600 ± 50
800 ± 150
1100 ± 150
1375 ± 125
Limits of permissible absolute error of measurement of hardness on the Vickers scale, HV, (±)
HV30, HV100
± 3
±5
± 7
± 8
±10
± 12
± 16
± 24
± 30
 
Limits of permissible absolute error of measurement of hardness instrument versatile instrument for measuring the hardness method Rockwell, Brinell, Vickers UT 5011A for Brinell listed in Table 5;
 
Table 5: Maximum permissible absolute error of hardness measurement device universal instrument for measuring the hardness of the method of Rockwell, Brinell, Vickers UT 5011A for Brinell
Marking scales Hardness
Measurement intervals Brinell hardness, HBW
40 ± 10
100 ± 50
200 ± 50
300 ± 50
400 ± 50
525 ± 75
Limits of absolute error by Brinell, HBW, (±)
HBW 2,5 / 62,5;
±2
± 6
±10
 
 
 
HBW 2,5 / 187,5
 
± 6
±10
± 12
± 18
± 24
 
The distance from the tip to the top of the test work plane device universal instrument for measuring the hardness of the method of Rockwell, Brinell, Vickers UT 5011A adjustable:
- Using a sliding table mounted on the PS1 leveling screw PV1 - from 0 mm to 140 mm;
- Without the use of the sliding table - from 0 mm to 240 mm;
Distance from the tip to the wall of the test enclosure universal instrument for measuring the hardness of the method of Rockwell, Brinell, Vickers UT 5011A, which limits the size of the test items - 145mm;
The hydraulic system provides smooth application base load for 2c-8c;
Instrument versatile instrument for measuring the hardness method Rockwell, Brinell, Vickers UT 5011A has a digital electronic control unit;;
Power - 220V, 50Hz;
Power consumption of the device versatile instrument for measuring the hardness method Rockwell, Brinell, Vickers UT 5011A - no more than 10 W;
Overall dimensions:
- Length - 490mm;
- Width - 170mm;
- Height - 780mm;
Weight of the device versatile instrument for measuring the hardness method Rockwell, Brinell, Vickers UT 5011A - 95kg;
Full average life of the device - not less than 10 years;
Instrument versatile instrument for measuring the hardness method Rockwell, Brinell, Vickers UT 5011A allows reference to evaluate the hardness of HRC in the range according to the values ​​shown in Table 6;
 
Table 6 Supplemental estimate hardness instrument versatile instrument for measuring the hardness method Rockwell, Brinell, Vickers UT 5011A HRC
Hardness scale
Load, N
View of the indenter
Measuring range, HR
D
980.7 N
Diamond cone
from 40 to 77
E
980.7 N
Ball Ø3, 175mm
70 to 100
F
588.4 N
Ball Ø1, 588mm
60 to 100
G
1471N
Ball Ø1, 588mm
from 30 to 94
H
588.4 N
Ball Ø3, 175mm
30 to 100
K
1471N
Ball Ø3, 175mm
40 to 100
L
588.4 N
Ball Ø6, 35mm
from 20 to 115
M
980.7 N
Ball Ø6, 35mm
from 20 to 115
P
1471N
Ball Ø6, 35mm
20 to 100
R
588.4 N
Ball Ø12, 7mm
from 20 to 115
S
980.7 N
Ball Ø12, 7mm
20 to 100
V
1471N
Ball Ø12, 7mm
20 to 100
 
Instrument versatile instrument for measuring the hardness method Rockwell, Brinell, Vickers UT 5011A allows reference to evaluate hardness Brinell in line with the values ​​specified in Table 7.
 
Table 7 Background score hardness instrument versatile instrument for measuring the hardness method Rockwell, Brinell, Vickers UT 5011A for Brinell
Hardness scale
Load, N
View of the indenter
Measuring range, NBW
HBW 5/62, 5
612.9 N
Ball Ø5mm
8-50
HBW 10/100
980.7 N
Ø10mm ball
3-20
HBW 1/30
294.2 N
Ball Ø1mm
95-450
 
The device allows to evaluate the reference hardness of Vickers, HV:
- In the range of up to 14HV 50HV;
- In the range of up to 1500HV 3000HV;
Completeness of the delivery device universal instrument for measuring the hardness method Rockwell, Brinell, Vickers UT 5011A provided to the extent necessary for its installation and commissioning in accordance with the documentation and working conditions of the order;
The package includes product maintenance documentation (User Manual);
Completeness of the delivery device universal instrument for measuring the hardness method Rockwell, Brinell, Vickers UT 5011A is provided in accordance Table 8;
 
Table 8: Delivery device universal instrument for measuring the hardness of the method of Rockwell, Brinell, Vickers UT 5011A
Document Designation
The name and symbol
Quantity
Note
UT5011AA
instrument
1 pc.
 
TA1
Movable table
1 pc.
 
OP1
Optical system
1 pc.
 
Replacement parts
CT1
Flat table test Ø139mm
1 pc.
 
CT3
Prismatic Table Test Ø39mm 90 °
1 pc.
 
GOST 9377-81
Tax tip Rockwell
1 pc.
In the case
GOST 9377-81
Tax tip Vickers
1 pc.
In the case
 
Tip 1.588
1 pc.
 
 
Tip 2.5
1 pc.
 
OB5
5x lens
1 pc.
In the case
OB10
10x lens
1 pc.
In the case
Conditioning unit versatile instrument for measuring the hardness method Rockwell, Brinell, Vickers UT 5011A
GOST 9031-75
Hardness model 2nd class MTR-1
1 com.
In the case
GOST 9031-75
Hardness model 2nd class MTB-1
1 com.
In the case
GOST 9031-75
Hardness model 2nd class MTB (100 ± 25) HV30
1 pc.
 
GOST 9031-75
Hardness model 2nd class MTB (400 ± 75) HV30
1 pc.
 
GOST 9031-75
Hardness model 2nd class MTB (800 ± 50) HV30
1 pc.
 
GOST 9031-75
Hardness model 2nd class MTB (450 ± 75) HV100
1 pc.
 
 
Set of keys
1 pc.
 
Operational documentation
 
Operating manual for the device
1 pc.
 
 
Passport to model hardness
1 pc.
 
 
Verification certificate
1 pc.
 
 
The additional multi-purpose device with the apparatus for measuring the hardness method Rockwell, Brinell, Vickers UT 5011A (available upon request) under Table 9;
 
Table 9 The additional multi-purpose device with the apparatus for measuring the hardness method Rockwell, Brinell, Vickers UT 5011A
Document Designation
The name and symbol
Quantity
Note
PF1
Retaining clip
1 pc.
 
CT2
Flat table test Ø179mm
1 pc.
 
CT4
Prismatic Table Test Ø58mm angle of 150 °
1 pc.
 
 
Tip 5
1 pc.
 
 
Tip 3.175
1 pc.
 
 
Tip 6.35
1 pc.
 
 
Tip of 12.7
1 pc.
 
 
Tip 10
1 pc.
 
Replacement of the unit versatile instrument for measuring the hardness method Rockwell, Brinell, Vickers UT 5011A
GOST 3722-81
Balls (1,588 ± 0,003) mm.
5 pc
In the case
GOST 3722-81
Balls (2,5 ± 0,003) mm.
5 pc
In the case
GOST 3722-81
Balls (5 ± 0,003) mm.
5 pc
In the case
GOST 3722-81
Balls (3,175 ± 0,003) mm.
5 pc
In the case
GOST 3722-81
Balls (6,35 ± 0,003) mm.
5 pc
In the case
GOST 3722-81
Balls (12,7 ± 0,003) mm.
5 pc
In the case
GOST 3722-81
Balls (10 ± 0,003) mm.
5 pc
In the case
 
The manufacturer guarantees that the quality of the instrument versatile instrument for measuring the hardness method Rockwell, Brinell, Vickers UT 5011A technical requirements specifications provided the customer terms and conditions of storage, transportation, installation, operation, and its specified operational documentation.


Яндекс цитирования