Radiometric devices >> oscilloscopes >> universal oscilloscope

S1-108 (C1-108)

34 999 RUB
13 999 UAH
available:
5 pieces

Online prices include VAT.

For order is possible effectively delivery of any quantity.

Universal oscilloscope S1-108

Also, this device may be referred to: S1108, S1 108, c1-108, c1108, c1 108.

 

S1-108 universal oscilloscope is used to investigate the continuous and pulse shapes, including rarely repeated single signals and voltages in the range from 20 mV to 8 V and lengths ranging from 7 ns to 100 ms by visual observation and photography.

 

Specifications S1-108:

 

Accuracy class - 2.

Oscilloscope channels - 1.

Effective screen - 80×100 mm.

The line width of each beam - not more than 0,8 mm.

 

Recording speed (RF-3):

- sensitivity - 1200 I/X-ray;

- relative aperture of 1:1,5 lens - not less than 2000 km/s;

- the lens aperture with 1:2 - not less than 1200 km/s.

Coefficient of variation of channel S1-108 - 10 mV/div to 1 V/div (or series of numbers 1, 2, 5).

 

Error deflection factors:

- basic - no more than 2%;

- in the range of each influence quantity in the working conditions of use - no more than 2,5%.

 

Basic measurement error voltage - no more than (0,5×10 1/n)%,

where n - the size of the image of the measured voltage in scale divisions (in centimeters) of the oscilloscope screen:

- by direct inlet in the range from 20 mV to 8 V;

- active probe in the range from 20 V to 4 V;

- a remote 1:10 divider between 200 mV to 10 V.

 

Basic measurement error voltage - no more than (2,5+10/n)%:

- 1:50 with external divider between 500 mV to 25 V.

 

Voltage measurement error S1-108 in the range of each influence quantity in the working conditions of application of the direct input:

- with an active probe and external divider 1:10 - no more (1,2+0,5·10/n)%;

- with external divisors 1:50 - (3+10/n)%.

 

Surge characteristic oscilloscope

 

Rise time:

- direct input and external divider transmission factor 1:10, 1:50 - less than 1 ns;

- with an active probe - 1,5 ns.

 

Emissions S1-108:

- direct input and external divider 1:10 1:50 - no more than ±5%;

- with an active probe - 7%.

 

Settling time:

- with direct input - no more than 5 ns;

- with an active probe - not more than 8 ns;

- with external divider 1:10, 1:50 - no more than 20 ns.

 

Irregularity in the area of establishing:

- with direct input - 5%;

- with an active probe and external divider 1:10, 1:50 - 7%.

 

Uneven with direct input:

- in the area of up to 100 ns - 2%;

- in the area from 100 ns to 100 ms - 1,5%.

 

Uneven with an active probe and external divider 1:10, 1:50:

- in the area of up to 100 ns - 3%;

- in the area from 100 ns to 100 ms - 2%.

Distortion DC - no more than 2%.

 

The drift of convergence at the screen S1-108:

- long-term drift - not more than 5 mm/h;

- short-term drift - less than 2 mm.

Periodic and random deviations from the active probe - not more than 2 mV.

 

Offset ray oscilloscope:

- when the voltage supply by 10% - less than 2 mm;

- at an input current - no more than 3 mm.

The limits of vertical movement of the beam - at least ±80 mm.

 

Parameters input channel vertical deflection

 

With direct input:

- input impedance - 50±1 Ohm;

- reflectance - no more than 10%.

 

With external divider 1:10:

- input impedance - 500±5 Ohm;

- input capacity - not more than 1 pF;

 

With external divider 1:50:

- input resistance - 2500±25 Ohm;

- input capacity - not more than 1 pF;

 

With an active probe:

- input impedance - 100±5 kOhm;

- input impedance divider 1:10 - 1±0,05 MOhm;

- input capacity - not more than 4 pF;

- input capacitance divider 1:10 - no more than 2,5 pF.

 

Picture delay signal S1-108 - not less than 12 ns.

Scan rate range (stepwise series of numbers 1, 2, 5 respectively) - from 1 ns/div to 10 ms/div.

 

Basic error coefficients sweep the work area sweep:

- from 10 ns/div to 10 ms/div - no more than 2%;

- from 1 ns / div to 5 ns/div - no more than 4%.

 

Voltage measurement error in the range of each influence quantity in the working conditions of use:

- from 10 ns/div to 10 ms/div - 2,5%;

- from 1 ns/div to 5 ns/div - 5%.

 

The basic error of S1-108 measuring time intervals within the working area sweep:

- from 10 ns/div to 10 ms/div - not more than (0,5×0,4·10/l)%,

where l - image size measured time interval in ticks (in centimeters) screen scale;

- from 1 ns/div to 5 ns/div - no more than (2+10/l)%.

 

Error of measurement of time intervals in the range of each influence quantity in the working conditions of use:

- from 10 ns/div to 10 ms/div - not more than (0,7×0,4·10/l)%;

- from 1 ns/div to 5 ns/div - no more than (2,5+10/l)%.

 

Frequency measurement error - no more than ±1 on the error measure time intervals.

Limits of movement of the horizontal beam - at least ±50 mm.

 

Internal synchronization:

 

Frequency range S1-108 of the harmonic signal - to 50 Hz; from 350 MHz.

 

Minimum level in the frequency range of the harmonic signal oscilloscopes:

- up to 50 Hz; from 100 MHz (pulse duration - 5 ns) - 8 mm;

- up to 100 MHz; from 350 MHz - 20 mm.

 

The maximum level of the frequency range of harmonic signals:

- up to 50 Hz; from 100 MHz (pulse duration - 5 ns) - 80 mm.

Instability sync - no more than 0,05+0,1/Csnom cases where

Csnom - nominal value of the set sweep rate.

 

External synchronization S1-108

 

Harmonic signal frequency range - up to 50 Hz; from 350 MHz.

 

Minimum level in the frequency range of the harmonic signal oscilloscope:

- up to 50 Hz; from 100 MHz (pulse duration - 5 ns) - 100 mV;

- up to 100 MHz; from 350 MHz - 200 mV.

 

The maximum level of the frequency range of harmonic signals:

- up to 50 Hz; from 100 MHz (pulse duration - 5 ns) - 3 V.

Instability sync - no more than 0,05+0,1/Csnom div.

 

Built-in calibrator

 

External load S1-108 - 50±1 Ohm.

 

Meander signal:

- repetition period - 10±0,01 ms;

- amplitude - 0,5±0,0025 B.

 

Harmonic signal:

- repetition period - 10±0,02 ns;

- amplitude - not less than 0,1 V.

 

Mode Tags

 

Number of measuring marks - 2 (reference and a reading).

 

Limit the scale of S1-108:

- vertical - at least ±40 mm;

- horizontally - of at least ±50 mm from the center of the screen.

 

Movement of the bearing with respect to the reference mark:

- vertically - from 73 mm to 80 mm;

- horizontally - from 95 mm to 100 mm.

Level accuracy of the test signal "0,1", "0,5", "0,9" - no more than 1% of the level set to "1".

 

Geometric distortion on the borders of the scale of the CRT screen:

- horizontal - not more than 3%;

- vertical - not more than 2%.

Orthogonality error S1-108 - not more than 1°.

 

Modes scans - autooscillation, waiting, single.

Calculation of percentages provided two units with digital display of results on the scoreboard.

Warm-up time - 15 minutes.

Dimensions - 526×200×304 mm.

Weight - no more than 17 kg.

 

Operating conditions S1-108:

 

Ambient temperature:

- operating - from 5° C to 40° C;

- limit - from -50º C to +60° C.

 

Relative humidity at 30° C:

- working - up to 85%;

- limit - up to 95%.

 

Mains voltage - 220±22 V.

Frequency - 50±0,5 Hz, 60 Hz.

 

Power consumption oscilloscopes:

- at rated voltage - no more than 110 V·A;

- with an active probe included - no more than 110 V·A.

 

MTBF device S1-108 - not less than 4000 hours.

Lifetime (80%) - 10 years.

Resource (80%) - 10 000 h.

 

Shelf life (50%):

- in a heated room - 7 years;

- in an unheated room - 5 years.

Photos S1-108

S1-108 (C1-108) device image.
S1-108 (C1-108) device image.
S1-108 (C1-108) scheme image.
S1-108 (C1-108) scheme image.