Radiometric devices >> oscilloscopes (S1-, S7-, S8-, S9-, Ya40-, Ya4S-, Ya9S- and others) >> universal oscilloscope (S1-, Ya40-, Ya4S- and others)
|
|||||||||||||||||||||||||
Universal oscilloscope S1-108 Also, this device may be referred to: S1108, S1 108, c1-108, c1108, c1 108.
S1-108 universal oscilloscope is used to investigate the continuous and pulse shapes, including rarely repeated single signals and voltages in the range from 20 mV to 8 V and lengths ranging from 7 ns to 100 ms by visual observation and photography.
Specifications S1-108:
Accuracy class - 2. Oscilloscope channels - 1. Effective screen - 80×100 mm. The line width of each beam - not more than 0,8 mm.
Recording speed (RF-3): - sensitivity - 1200 I/X-ray; - relative aperture of 1:1,5 lens - not less than 2000 km/s; - the lens aperture with 1:2 - not less than 1200 km/s. Coefficient of variation of channel S1-108 - 10 mV/div to 1 V/div (or series of numbers 1, 2, 5).
Error deflection factors: - basic - no more than 2%; - in the range of each influence quantity in the working conditions of use - no more than 2,5%.
Basic measurement error voltage - no more than (0,5×10 1/n)%, where n - the size of the image of the measured voltage in scale divisions (in centimeters) of the oscilloscope screen: - by direct inlet in the range from 20 mV to 8 V; - active probe in the range from 20 V to 4 V; - a remote 1:10 divider between 200 mV to 10 V.
Basic measurement error voltage - no more than (2,5+10/n)%: - 1:50 with external divider between 500 mV to 25 V.
Voltage measurement error S1-108 in the range of each influence quantity in the working conditions of application of the direct input: - with an active probe and external divider 1:10 - no more (1,2+0,5·10/n)%; - with external divisors 1:50 - (3+10/n)%.
Surge characteristic oscilloscope
Rise time: - direct input and external divider transmission factor 1:10, 1:50 - less than 1 ns; - with an active probe - 1,5 ns.
Emissions S1-108: - direct input and external divider 1:10 1:50 - no more than ±5%; - with an active probe - 7%.
Settling time: - with direct input - no more than 5 ns; - with an active probe - not more than 8 ns; - with external divider 1:10, 1:50 - no more than 20 ns.
Irregularity in the area of establishing: - with direct input - 5%; - with an active probe and external divider 1:10, 1:50 - 7%.
Uneven with direct input: - in the area of up to 100 ns - 2%; - in the area from 100 ns to 100 ms - 1,5%.
Uneven with an active probe and external divider 1:10, 1:50: - in the area of up to 100 ns - 3%; - in the area from 100 ns to 100 ms - 2%. Distortion DC - no more than 2%.
The drift of convergence at the screen S1-108: - long-term drift - not more than 5 mm/h; - short-term drift - less than 2 mm. Periodic and random deviations from the active probe - not more than 2 mV.
Offset ray oscilloscope: - when the voltage supply by 10% - less than 2 mm; - at an input current - no more than 3 mm. The limits of vertical movement of the beam - at least ±80 mm.
Parameters input channel vertical deflection
With direct input: - input impedance - 50±1 Ohm; - reflectance - no more than 10%.
With external divider 1:10: - input impedance - 500±5 Ohm; - input capacity - not more than 1 pF;
With external divider 1:50: - input resistance - 2500±25 Ohm; - input capacity - not more than 1 pF;
With an active probe: - input impedance - 100±5 kOhm; - input impedance divider 1:10 - 1±0,05 MOhm; - input capacity - not more than 4 pF; - input capacitance divider 1:10 - no more than 2,5 pF.
Picture delay signal S1-108 - not less than 12 ns. Scan rate range (stepwise series of numbers 1, 2, 5 respectively) - from 1 ns/div to 10 ms/div.
Basic error coefficients sweep the work area sweep: - from 10 ns/div to 10 ms/div - no more than 2%; - from 1 ns / div to 5 ns/div - no more than 4%.
Voltage measurement error in the range of each influence quantity in the working conditions of use: - from 10 ns/div to 10 ms/div - 2,5%; - from 1 ns/div to 5 ns/div - 5%.
The basic error of S1-108 measuring time intervals within the working area sweep: - from 10 ns/div to 10 ms/div - not more than (0,5×0,4·10/l)%, where l - image size measured time interval in ticks (in centimeters) screen scale; - from 1 ns/div to 5 ns/div - no more than (2+10/l)%.
Error of measurement of time intervals in the range of each influence quantity in the working conditions of use: - from 10 ns/div to 10 ms/div - not more than (0,7×0,4·10/l)%; - from 1 ns/div to 5 ns/div - no more than (2,5+10/l)%.
Frequency measurement error - no more than ±1 on the error measure time intervals. Limits of movement of the horizontal beam - at least ±50 mm.
Internal synchronization:
Frequency range S1-108 of the harmonic signal - to 50 Hz; from 350 MHz.
Minimum level in the frequency range of the harmonic signal oscilloscopes: - up to 50 Hz; from 100 MHz (pulse duration - 5 ns) - 8 mm; - up to 100 MHz; from 350 MHz - 20 mm.
The maximum level of the frequency range of harmonic signals: - up to 50 Hz; from 100 MHz (pulse duration - 5 ns) - 80 mm. Instability sync - no more than 0,05+0,1/Csnom cases where Csnom - nominal value of the set sweep rate.
External synchronization S1-108
Harmonic signal frequency range - up to 50 Hz; from 350 MHz.
Minimum level in the frequency range of the harmonic signal oscilloscope: - up to 50 Hz; from 100 MHz (pulse duration - 5 ns) - 100 mV; - up to 100 MHz; from 350 MHz - 200 mV.
The maximum level of the frequency range of harmonic signals: - up to 50 Hz; from 100 MHz (pulse duration - 5 ns) - 3 V. Instability sync - no more than 0,05+0,1/Csnom div.
Built-in calibrator
External load S1-108 - 50±1 Ohm.
Meander signal: - repetition period - 10±0,01 ms; - amplitude - 0,5±0,0025 B.
Harmonic signal: - repetition period - 10±0,02 ns; - amplitude - not less than 0,1 V.
Mode Tags
Number of measuring marks - 2 (reference and a reading).
Limit the scale of S1-108: - vertical - at least ±40 mm; - horizontally - of at least ±50 mm from the center of the screen.
Movement of the bearing with respect to the reference mark: - vertically - from 73 mm to 80 mm; - horizontally - from 95 mm to 100 mm. Level accuracy of the test signal "0,1", "0,5", "0,9" - no more than 1% of the level set to "1".
Geometric distortion on the borders of the scale of the CRT screen: - horizontal - not more than 3%; - vertical - not more than 2%. Orthogonality error S1-108 - not more than 1°.
Modes scans - autooscillation, waiting, single. Calculation of percentages provided two units with digital display of results on the scoreboard. Warm-up time - 15 minutes. Dimensions - 526×200×304 mm. Weight - no more than 17 kg.
Operating conditions S1-108:
Ambient temperature: - operating - from 5° C to 40° C; - limit - from -50º C to +60° C.
Relative humidity at 30° C: - working - up to 85%; - limit - up to 95%.
Mains voltage - 220±22 V. Frequency - 50±0,5 Hz, 60 Hz.
Power consumption oscilloscopes: - at rated voltage - no more than 110 V·A; - with an active probe included - no more than 110 V·A.
MTBF device S1-108 - not less than 4000 hours. Lifetime (80%) - 10 years. Resource (80%) - 10 000 h.
Shelf life (50%): - in a heated room - 7 years; - in an unheated room - 5 years. Photos: S1-108 (C1-108) |