R4833 (P4833) R4833 (P4833) The device R4833
4.9 63
Device R4833 (R 4833, R-4833, p4833, p-4833, p 4833)
This device is designed to measure resistance, constant e.m.f. , voltage and calibration of thermotechnical devices.
Measuring range: bridge - from 10-4 Ohm to 106 Ohm, potentiometer - from 0 to 111.10 mV, resistance box from initial (≤ 0.015 Ohm) to 1111.10 Ohm.
Accuracy class: DC box 0.1; DC potentiometer 0.05; ; resistance box 0.02 / 1.5 • 10-4
 
Electrical measuring instruments
measures and boxes
R4833 (P4833) device image.

Price:

on request

 

Product information updated 29 September 2024.

Found cheaper?

Let us know. Our price is the best. 

Warranty:

12 months.

Warranty extension up to 5 years. 

Terms of delivery:

To all regions of the country, international delivery. 

print

Product information updated 29 September 2024.

Found cheaper?

Let us know. Our price is the best. 

Warranty:

12 months.

Warranty extension up to 5 years. 

Terms of delivery:

To all regions of the country, international delivery. 

print

Technical characteristics

Device R4833 (R 4833, R-4833, p4833, p-4833, p 4833)
This device is designed to measure resistance, constant e.m.f. , voltage and calibration of thermotechnical devices.
Measuring range: bridge - from 10-4 Ohm to 106 Ohm, potentiometer - from 0 to 111.10 mV, resistance box from initial (≤ 0.015 Ohm) to 1111.10 Ohm.
Accuracy class: DC box 0.1; DC potentiometer 0.05; ; resistance box 0.02 / 1.5 • 10-4

Devices R4833 (R-4833)
Device R-4833 multimeter (hereinafter - device) is designed to measure resistance, constant e.m.f. , voltage and calibration of thermotechnical devices.
     Devices R4833 working conditions: ambient temperature from 10 ° to 35 ° C, relative humidity 80% at 25 ° C, atmosphere pressure (84-106.7) kPa [(630-800) mmHg].
 
Technical characteristics
Devices R4833 accuracy class, when used as: dc bridge 0.1; dc potentiometer 0.05; resistance box 0.02/1.5•10-4 .
Range: resistance measurements, when using device as bridge from 10-4 to 106 Ohm;
     measuring e.m.f and voltage, when using device R-4833 as potentiometer from 0 to 111.10 mV (contour I -10 decades at 10 mV and 10 decades at 1 mV; contour II -10 decades at 0.1mV, and 10 decades mV at 0.01 mV);
resistance readings when using device as five-decaderesistance box from initial (≤ 0.015 Ohm) to 1111.10 Ohm (decade 100; 10; 1; 0.1; 0.01 Ohm).
 
Table 1
Measuredresistance, Ohm Backs ratio, N Voltage supply, V Additional resistance of external batteries, Ohm
External Internal circuits
10-4-10-3 10-4 - - 1.5 -
10-3-10-2 10-4 - - 1.5 -
10-2-10-1 10-3 1.5 1.5 1.5 3
10-1-1 10-2 1.5 1.5 1.5 3
1-5 10-2 1.5 1.5 1.5 3
5-102 10-1 1.5 1.5 1.5 3
102-103 1 7 7 36 -
103-104 10 7 7 36 -
104-105 102 48 7 36 510
105-106 103 48 7 36 510
 
Measuring resistance, Ohm Measurement error,%, not more than Parameters of external galvanometer
Internal galvanometer(IG) External galvanometer (EG)- any power supply
Internal battery(IB) Bridges battery external(BB) circuits
10-4-10-3 - ±5.0 - ±0.5* Rr≤20 Ом
10-3-10-2 - ±1.0 ±2.0 ±0.5* Ci≤ 1· 10-7 А/div
10-2-10-1 ±0.5 ±0.5 ±0.5 ±0.1* Ci≤ 5· 10-7 А/div
10-1-1 ±0.2* ±0.2* ±0.2* ±0.1* Rr≤20 Ом
1-5 ±0.1 ±0.1 ±0.1 ±0.1  
5-102 ±0.1 ±0.1 ±0.1 ±0.1  
102-103 ±0.1 ±0.1 ±0.1 ±0.1  
103-104 ±0.1 ±0.1 ±0.1 ±0.1  
104-105 ±1.0 ±0.5 ±0.5 ±0.1 Ci≤2 · 10-8 А/div
105-106 - ±2.0 ±5.0 ±0.1 Rr≤20 Ом
 
Note. When measuring resistances in range (10-4-10-3)Ohm ; (10-3-10-2) Ohm external current power supply should be 3A. Significance of back comparison initial resistance should be considered.
   Basic error:
1) When using device R4833 as bridge, the acceptable basic error limit (Δ) in Ohm corresponds to value defined by formulas:
    ∆ = ±10-2 · С · (10-1Rn + X)(1)
   or
    ∆= ±10-2 ·С · X, (2)
where C - accuracy class value, specified for given range (Table 1); Rn - normalizing resistance value for given measurement range, Ohm (Table 1); X - reading, measured from decades switches limbs, Ohm.
     Acceptable basic error limit corresponds to values specified in Table 1, under normal using conditions:
ambient temperature (20 ± 5) ° C; relative humidity of 25 to 80%; atmosphere pressure (84 - 106.7) kPa [(630-800) mm Hg];
working position - horizontal;
2) When using device R-4833 as potentiometer for emf and voltage acceptable basic error limit measurements (ΔU) in V corresponds to values defined by formula:
∆U= ±5 · 10-4 · (Un/10 + U),(3)
where Un - normalizing value ,V (Un = 0.1 V);
U- potentiometer reading, V, under normal using conditions:
ambient temperature (20 ± 2) ° C;
relative humidity of 25 to 80%;
atmosphere pressure (84-106.7) kPa [(630-800) mm Hg];
working position - horizontal;
no external magnetic field, except the earth one.
     When using device R4833 as potentiometer it must meet requirements of  basic error (instability) limits in accordance with accuracy class during one year from the date of first certification (calibration) under normal conditions (above) and under normal conditions of use, transportation and storage.
3) Acceptable deviation limit of actual resistance value (δ) in percentage from nominal when using device as resistance box corresponds to value defined by formula:
    δ = ± [0.02+1.5 • 10-4 (Rk/R-1)], (4)
where  Rk - maximum resistance value, Ohm (Rk = 1111.10 Ohm); R-nominal resistance value, Ohm.
Actual resistance value deviation from nominal in primary calibration (at manufacturer release) and acceptable resistance change δ in percentage per year (instability) does not exceed value determined by formula (4) under normal using conditions:
ambient temperature (20 ± 2) ° C;
relative humidity of 25 to 80%;
atmosphere pressure (84-106.7) kPa [(630-800) mm Hg.];
working position - any;
steady-state thermal equilibrium and power dissipation are not above nominal.
Acceptable basic error of backs resistances ratio is not more than  ± 0.025% (except the level with nominal value 0.09864 Ohm, for which acceptable error is not more than ± 0.25%) in normal using conditions (above).
Acceptable basic error of resistors box 2.5 and 7.5 Ohm is not more than ± 0.1%.
Built-in R4833 multimeter device regulated voltage source IPH ("mV") at voltage supply at least 1.3 V and load resistance not less than 25 Ohm provided at clamps
«-X», «mV»  voltage from minus 5V to plus 100 mV.
Voltage regulation discreteness is not more than 0.05% from maximum voltage at clamps «-X», «mV».
DC Bridge  built into device R-4833 is regulated power source IPH ("V") at voltage supply at least 5.6 V provides voltage from 0.5 to 5V at clamps «-X», «B».
Comparison resistors acceptable basic error in circuit with measuring connecting lines resistance for automatic bridges and ratio-meter is not more than ± 0.02%.
    Primary devices voltage U0 when used as potentiometer is 2.4·10-6V, resolution ΔUn is not more than 25·10-6V. Scale interval of smallest decade on one stage Umin= 10-5V.
Potentiometers installation resistance regulated part provides changes in installation steps, which do not exceed 0.01% from emf normal element value (1.0184-1.0194) V.
Primery resistance average value R0 of comparison devices back, resistance during installation of all decadal switches on zero, is not more than 0.015 Ohm.
Primary resistance variation ΔR0 of comparison back, caused by changes in contact resistance  switching devices, is not more than 0.0015 Ohm.
Nominal power dissipation on one decade  is not more than: 0.001 Watt for 0.01 Ohm, 0.01 Watt for 0.1 Ohm, 0.05 Watt for1 Ohm  and above.
Maximum power dissipation by one decade  is not more than: 0.005 Watt for 0.01 Ohm, 0.05 Watt for 0.1 Ohm, 0.1 Watt for 1 Ohm and higher.
 
Additional error.
When using device R4833 as bridge additional error limit in working temperature ranges does not exceed maximum acceptable basic error when ambient temperature changes for every 10 ° C.
When using device as potentiometer changes in readings, shown in percentage from  basic error does not exceed 100% when ambient temperature changes for every 10 ° C in range of working temperatures.
Note. Additional error caused by operating current installation is  not specified, because it is in basic error.
When using device as resistance box additional error limit in working temperature range of 15 to 25 ° C, caused by changes in ambient temperature (environment) between upper (lower) limit of normal temperature conditions range 22 ° C (18 ° C) and some point in related working conditions temperature range 25 ° C (15 ° C), corresponds to the largest change in resistance Rmax, within working temperature from 10 to 15 ° C and 25 ° to 35 ° C for every 5 ° C of temperature change  do not exceed value determined by  formula (4).
When using device R-4833 as resistance box additional error limit as a percentage of its nominal value when power dissipation change from normal to any value do not exceed maximum power under normal using conditions and steady-state thermal equilibrium do not exceed value determined by the formula (4).
   
Power supply.
When used as a bridge - from built-in galvanic elements with voltage at least 1.5 mV (with four clamping connection of measured resistance from 10-4 to 102 Ohm), with voltage not less than 7.0 V (with two clamping connection of measured resistances from 102 to 106  of permissible DC pulsation, power supplies from 0 to 0.1%, maximum power consumed do not exceed 3A and 0.05 A; from external power supply according to table 1; from AC voltage (220 ± 22)V, frequency (50 ± 0.5) or (60 ± 0.6) Hz, sinusoidal waveform, the higher harmonics coefficient is not more than 5%.
When used as potentiometer - from built-in and external sources of DC voltage from 1.20 V to 1.65 V.
Power IPH ("mV") - from built-in and external dc sources with voltage from 1.20 to 1.65 V.
     Power IPH ("V") - from built-in and external dc sources with voltage from 4.8 to 6.6 V. Acceptable dc pulsation source from 0 to 0.7%, maximum power consumption is not more than 0.025 A.
Power consumed by device bridge R-4833 does not exceed 10VA.
Devices operating mode time does not exceed 15 minutes, after turn on. Working mode - intermittent.
Devices continuous operation time from system and built-in power supply is at least 8 hours. Interval time before re-enable is at least 15 minutes.
Insulation resistance between case and isolated dc electric circuits in working conditions is not less than 5·1011 Ohm at dc voltage (500 ± 100) V.
 
Insulation between case and insulated from case dc electrical circuits (all interconnected clamps) can withstand for 1 min test voltage of 0.5 kV, AC frequency 50 Hz in working condition, and between plug and case - 1.5 kV.
Resistance of each calibrated cables used in resistance measurement is from 0.0012 to 0.0015 Ohm.
Resistance of each calibrated cables, used for ratiometers and bridges verification is from 0.027 to 0.033 Ohm.
Resistance error (0.6; 1.6; 5; 15; 16.2; 25 Ohm) for simulation of  trunk pyrometric millivoltmeters is not more than ± 0.1 Ohm.
Dimensions are not more than 250x390x190 mm. Bridge R-4833 (R4833) weight is not more: without packaging - 8 kg; in transport container - 20kg. Total content of nonferrous metals, kg: aluminum D16A - 0.7; copper and copper-based alloys: copper wire MM - 0.3; bronze BrOF 6.5-0.15-1 - 0.2, brass L63 - 1.02; brass LC59-1 - 0.5.
Rate the product: R4833 (P4833)
Number of views: 2367
Product rating: 4.5
Number of grades: 431
* rating and all product statistics are updated once a month; information is current as of 16 October 2024

Photos: R4833 (P4833)

R4833 (P4833) device image.

R4833 (P4833) device image.

R4833 (P4833) view panel of the device.

R4833 (P4833) view panel of the device.

R4833 (P4833) front view.

R4833 (P4833) front view.

R4833 (P4833) side view.

R4833 (P4833) side view.

R4833 (P4833) rear view.

R4833 (P4833) rear view.

R4833 (P4833) bottom view.

R4833 (P4833) bottom view.

Leading online store Zapadpribor - a huge selection of measuring equipment for the best value and quality. So you can buy inexpensive devices, we monitor competitors' prices and are always ready to offer a lower price. We only sell quality products at the best prices. On our site you can buy cheaply as the latest innovations and proven equipment from the best manufacturers.

The site includes a special offer «buy at the best price» - if other Internet resources (bulletin board, forum or announcement of another online service) in products presented on our site, the lower price, we will sell it to you even cheaper! Buyers also additional discount for leaving a review or photo of use of our products.

In the price list contains not the whole range of products offered. Prices of goods, not included in the price list can learn to contact the manager. Also, our managers You can get detailed information about how cheap and profitable to buy instrumentation wholesale and retail. Telephone and e-mail for advice on the purchase , delivery or receipt are shown on the description of the goods. We have the most qualified staff, quality equipment and the best price.

 

Online store Zapadpribor - the official manufacturer's dealer test equipment. Our goal - selling high quality goods with best price offer and service to our customers. Our store can not only sell you the necessary instrument, but also offer additional services to its calibration, repair and installation. That you have a pleasant experience when shopping on our site, we have provided special gifts guaranteed to the most popular products.

 

We offer fast international shipping to almost every country in the world: Australia, Austria, Azerbaijan, Albania, Algeria, Anguilla, Angola, Antigua and Barbuda, Argentina, Aruba,  Bahamas, Bangladesh, Barbados, Bahrain, Belize, Belgium, Benin, Bermuda, Bulgaria, Bolivia, Bonaire, Sint-E. and Saba, Bosnia and Herzegovina, Botswana, Brazil, British Virgin Islands, Brunei Darussalam, Burkina Faso, Burundi, Bhutan, VietNam, Vanuatu, Vatican City, Venezuela, Armenia, Gabon, Guyana, Haiti, Gamia, Gambia, Ghana, Guatemala, Guinea, Gibraltar, Honduras, Hong Kong, Grenada, Greenland, Greece, Georgia, Denmark, Democratic Republic of the Congo, Jersey, Djibouti, Dominica, Dominican Republic, Ecuador, Eswatine, Estonia, Ethiopia, Egypt, Zambia, Zimbabwe, Jordan Indonesia, Ireland, Iceland, Spain, Italy, Cape Verde, Kazakhstan, Cayman Islands, Cambodia, Cameroon, Canada, Qatar, Kenya, Kyrgyzstan, China, Cyprus, Kiribati, Colombia, Comoros, Congo, Korea (Rep.), Costa Rica, Kot-d'Ivoire, Cuba, Kuwait, Curacao, Laos, Latvia, Lesotho, Lithuania, Liberia, Lebanon, Libya, Liechtenstein, Luxembourg, Myanmar, Mauritius, Mauritania, Madagascar, Macao, Malawi, Malaysia, Mali, Maldives, Malta, Morocco, Mexico, Mozambique, Moldova, Monaco, Monaco, Namibia, Nauru, Nepal, Niger, Nigeria, Netherlands, Germany, New Zealand, New Caledonia, Norway, UAE, Oman, Cook Islands, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, South Africa, Poland, Portugal, Republic of Chad, Rwanda, Romania, El Salvador, Samoa, San Marino, Saudi Arabia, Swaziland, Seychelles, Senegal, Saint Vincent and the Grenadines, Saint Kitts and Nevis, Saint Lucia, Serbia, Singapore, Sint Maarten, Slovakia, Slovenia, Solomon Islands, United Kingdom of Great Britain and Northern Ireland, Sudan, Suriname, East Timor (Timor-Leste), USA, Sierra Leone, Tajikistan, Taiwan, Thailand, Tanzania (United Republic of), Togo, Tonga, Trinidad and Tobago, Tuvalu, Tunisia, Turkey, Turkmenistan, Uganda, Hungary, Uzbekistan, Uruguay, Faroe Islands, Fiji, Philippines, Finland, France, French Polynesia, Croatia, Central African Republic, Czech Republic, Chile, Montenegro, Switzerland, Sweden, Sri Lanka, Jamaica, Japan.

We also provide such metrological procedure: calibration, tare, graduated, testing of measuring instruments.

The online store takes an active part in such procedures as electronic bidding, tender, auction.

If there is no site in the descriptive information you need on the device you can always turn to us for help. Our qualified managers will update for you on the technical characteristics of the device from its technical documentation: user manual, certificate, form, operating instructions, diagram. If necessary, we will take a photo of your device, or a device stand. You can leave feedback on the unit purchased from us, the meter, the device, indicator, or product. Your review for your approval will be published on the site without contact information.

 

Description of the instruments taken from the technical documentation or technical literature. Most pictures of products made directly by our experts prior to shipment of goods. In the description of the device provided the main technical characteristics of the instrument: nominal measuring range, accuracy class, scale, supply voltage, size (size), weight. If the site you see the discrepancy device name (model) specifications, photo or documents attached - let us know - you will get a useful gift with-selling device.

 

While the need to clarify the overall weight and size or the size of the individual meter you can in our service center. If you need our engineers will help you choose the most complete analog or a suitable replacement for the device you are interested. All counterparts and replacement will be tested in one of our laboratories with the full compliance to your requirements.

In the technical documentation for each device or product, information is provided on the list and amount of precious metals content. The documentation gives the exact weight in grams of precious metals: Au gold, palladium Pd, platinum Pt, silver Ag, tantalum Ta and other platinum group metals (PGM) per unit item. These precious metals are found in nature in very limited quantities and therefore have such a high price.

On our website you can familiarize yourself with the technical characteristics of devices and obtain information on the content of precious metals in devices and radio components manufactured in the USSR. We draw your attention to the fact that often the actual content of precious metals differs by 10-25% from the reference one in a smaller direction! The price of precious metals will depend on their value and mass in grams.

All text and graphic information on the site is informative. The color, shade, material, geometric dimensions, weight, content, delivery set and other parameters of the goods presented on the site may vary depending on the batch of production and year of manufacture. Check with the sales department for more information.

 

If you can make repairs to the device yourself, our engineers can provide you with a complete set of necessary technical documentation: circuit diagram TO, ER, FD, PS. We also have an extensive database of technical and metrological documents: technical specifications (TS), the terms of reference (TOR), GOST (DSTU), the industry standard (OST) test procedure, the method of certification, verification scheme for more than 3,500 types of measuring equipment from the manufacturer this equipment. From the site you can download all the necessary software (the software driver) is required for the product purchased.

We carries out repairs and maintenance of measuring equipment in more than 75 different factories of the former Soviet Union and CIS.

We also have a library of legal documents that are related to our field of activity: the law, code, ordinance, decree, temporary position.