Voltage converter V9-6 (V96, V9 6, b96, b9-6, b9 6)

V9-6 measurement and conversion rms AC voltage of an arbitrary shape into a DC voltage in linear and logarithmic scale. Can be used together with any digital DC voltmeter.
Suppression is the DC 70 dB, with AC - 60 dB.
The range of measured and converted voltage 0.01V-300V.
The range of frequencies in the conversion mode - without the constant component of 10Hz-10MHz, with the DC - DC and 2 Hz - 10 MHz.
Tolerance conversion unit voltage converter V9-6 (percent) ± (0,3 + 0,2 Uc / U) - in the frequency range 20Hz-100kHz; ± (0,5 + 0,2 Uc / U) - in the frequency range 10Hz-20Hz; ± (O 6 + 0.4 Uc / U) - in the frequency range of 0.1 MHz to 1 MHz; ± (1 + 0.5 Uc / U) - in the frequency range of 5 MHz-1 ± (4 + Uc / U) in the frequency range 5-10 MHz; ± (1 + Uc / U) and the DC frequency range of 2-10Hz.
Internal conversion error logarithmic converter ± 0,25 dB.
Crest factor converted voltage at values ​​corresponding to finite values ​​of the sub-bands, 7.
The output voltage of the device voltage converter V9-6 when the signal corresponding to the final value of sub-band, 1 W.
Conversion time of 1s and 10s. Input impedance of 2 MW, the input capacitance of 40 pF. Dimensions 480x80x475 mm, weight 10 kg.

Photos: V9-6