Voltmeter V3-49 (V3 49, V349, b3-49, b3 49, b349)
Designed to measure the rms AC voltage sine wave is used for verification and calibration of electronic voltmeters, calibration standard signal generator output voltage level, and as a working instrument
Measuring range 10mV - 100V.
Error of 0.2%.

Voltmeter V3-49 AC compensation diode is designed to measure the rms AC voltage sine wave and is used for checking and calibrating electronic voltmeters calibration standard signal generator output voltage level, and also as a working tool.

 

Normal operating conditions:
- Air temperature 293 ± 5 K (20 ± 5 ° C);
- Relative humidity 65 ± 15%;
- Atmospheric pressure 100 ± 4 kPa (750 ± 30 mm Hg. Centuries).
- Mains voltage 220 ​​V ± 4,4 Operating conditions:
- Air temperature from 278 K (5 ° C) to 313 K (40 ° C);
- Relative humidity 90% at a temperature of 303 K (30 ° C);
- Atmospheric pressure 100 ± 4 kPa (750 ± 30 mm Hg. Centuries).
- Mains voltage of 220 ± 22 V, ± 0,5 at 50 Hz and harmonic content up to 5% and 220 ± 11 at 400 +28 / -12 Hz and harmonic content up to 5%.
Technical data
Measurement range of the instrument voltmeter V3-49 voltages from 10 mV to 100 V, four overlapping sub-bands with a resolution of 10 -4 of the final value sub-band:
a) from 10.00 to 100.00 mV mV
b) from 0.1000 V to 1.0000 in
c) from 1,000 V to 10,000
g) from 10.00 to 100.00 in In
The normal frequency range of the measured AC voltage of 20 Hz to 10 MHz, work - over 10 MHz to 1000 MHz. The basic error of the instrument in the normal range of frequencies, expressed as a percentage of reading does not exceed:
DOS δ = ± (0,2 + 0,08 / U)% (1)
where U - an instrument reading in volts.
The standard deviation of a random composition of Barking error (S) does not exceed one tenth of the basic error.
Error of the instrument voltmeter V3-49 working in the field of frequency tion is expressed as a percentage of the readings relative to the averaged frequency response shall not exceed:
δ = ± (0,2 + 0,08 / U + 0,008 f)% (2)
where f - frequency voltage measured in MHz. When an individual certification of measuring diodes instrument error in the working frequency range on the individual frequency response does not exceed
δ = ± (0,2 + 0,08 / U + 0,0035 f)% (3)
It shall be accompanied by a certified diodes individual characteristic of the frequency error. Change readings caused ambient air temperature deviation from normal to any temperature within the operating temperature range from 278 K (5 ° C) to 313 K (40 ° C) does not exceed one half of the basic error for every 10 K temperature change.
Change the voltmeter V3-49 readings caused by voltage tolerance at 50 ± 0,5 Hz, the nominal value of ± 22 V and the deviation of the supply voltage frequency of 400 +28 / -12 Hz from the nominal value by ± 11 V, does not exceed half of the basic error. In this case, the correction may be zero, and the filament voltage controls.
Active impedance device, measured at a frequency of 100 MHz is not less than 80 ohms. Input capacitance of the device is less than 1.5 pF. Standing wave ratio (K.S.V.N.) with the connected probe and loads agreed E9-001 E9-002 and not more than the values ​​specified in the table. 1.
 
Table 1
Connector Frequency MHz
300 700 1000
Standing wave ratio
P-010 1.1 1.3 1.5
P-011 1.1 1.3 1.5

Insulation resistance of the device voltmeter V3-49 at least: a housing unit - measuring chain - 1000 MW; screen - housing unit - 1 MW.
Voltmeter V3-49 provide their specifications within the standards set specifications, after samoprogreva time of 30 minutes.
The device retains its specifications within the standards set specifications, when powered from the mains AC voltage of 220 ± 22 V, 50 Hz ± 0,5 and harmonic content up to 5% and the mains voltage of 220 ± 11V at 400 +28 / - 12 Hz and harmonic content to 5%.
The power consumed by the device from the network at the rated voltage does not exceed 65 VA. The device provides the ability to remotely control the compensation voltage and measuring sub-bands, as well as remote control of zero-balance indicator.
Note. Zeroing filament voltage diode remotely is not available.
Voltmeter V3-49 has tsifropechatayuschee output to an external unit (CPU) in the code 8-4-2-1. The device issues a command to start to start the CPU after the measurement (obtaining a zero-balance indicator). The device allows for continuous operation in the operating conditions for 8 hours while maintaining its specifications within the standards set specifications. At the same time ensures the normal patterns of EEC, the RFP details and elements within the norms, standards and specifications for them.
Note. Continuous operation time does not include time samoprogreva.
Voltmeter V3-49 retain their specifications within the standards set by TU, after the change in the measurement of n diode 6D24N of SPTA.
Diodes complete spare parts of each device selected in the parameter "K" of the diode. Modes of operation of circuit elements meet the standards set out in the standards and specifications for them.
Life more than 5400 hours service life of 5 years. Lifespan 15,000 hours Unit Dimensions 90h175h475 4 mm. Dimensions of packing box with receipt of the customer 740h373h616 mm. Dimensions of packing box with receipt of OTC 612h195h515 mm. Overall dimensions of shipping containers 1064h562h768 mm. Unit weight voltmeter V3-49 to 20 kg. Weight with packaging not exceeding 50 kg. Weight with transport container not exceeding 100 kg.