AC Bridge R5083 (R-5083, R 5083, p5083, p-5083, p 5083)
 
AC Bridge R5083 is an autonomous measuring device for general industrial purposes and designed for automatic measurements of: capacitance C, inductance L, active resistance R, tangent loss tgδ (Q Qr), tangent angle of phase shift tgφ (Q Qc and QL) of measuremed objects, and also a percentage deviation of measured object with further presentation of measurement results in digital form.
AC bridge R5083 can be used for:
- Scientific researches;
- Metrological measuring instruments attestation;
- Quality control of electrical and radiotechnical products;
- Nonelectrical values measurement with a use of nonelectrical values transformers  into one of measured values.
 
Technical characteristics of AC bridge R5083:

Depending on frequency and voltage of measurement object, values measured by the AC bridge correspond to those shown in Table 1;
 
   Table1
 
Main
Additional
 
main
 
derivative
C
Cp
Rp
tgδc(QR)
Cs
Rs
L
Lp
Rp
tgδL(QR)
Ls
Rs
R
Rp
Cp
tgδc(QC)
Lp
tgδL(QL)
Rs
Cs
tgδc(QC)
Ls
tgδL(QL)
 
Accuracy classes of AC bridge R5083 for main measured values ​​are shown in Table 2;
 
Table 2 Accuracy Classes of AC for main measured values
   
 
Measured
Value
ƒon, ƒn
Measurement range
Accuracy class
number
counting
efficient
Cp, F;
Cs, F;
tgδC (QR)
0.999kHz, 0.100 kHz -0.999 kHz
1
000.01-099.99
10-12
0.2/0.02
100.000-999.99
0.1/0.01
2
1.00000-9.9999
10-9
0.05/0.005
3
10.00000-99.9999
4
100.0000-999.99
5
1.00000-9.9999
10-6
0.1/0.01
6
10.0000-99.999
0.2/0.02
7
100.000-999.99
0.5/0.2
8
1.00000-9.9999
10-3
1.5/0.15
9
10.000-9999
2.5/0.1
Cp. F;
Cs. F;
tgδC (QR)
1.00 kHz. 1.00 kHz -9.99 kHz
1
00.001-09.999
10-12
0.05/0.01
10.00000-99.9999
0.05/0.005
2
100.0000-999.99
0.02/0.002
3
1.00000-9.9999
10-9
4
10.0000-99.999
5
100.000-999.99
6
1.00000-9.9999
10-6
0.05/0.005
7
10.0000-99.999
0.1/0.01
8
100.000-999.99
0.5/0.05
9
1.00000-999.9
10-3
1/0.1
Cp. F;
Cs. F;
tgδC (QR)
10.0 kHz. 10.0 kHz -100 kHz
1
0.0001-0.9999
10-12
0.5/0.01
1.00000-9.9999
0.2/0.02
2
10.0000-99.999
0.1/0.01
3
100.000-999.99
4
1.00000-9.9999
10-9
5
10.0000-99.999
6
100.000-999.99
0.2/0.02
7
1.00000-9.9999
10-6
0.5/0.05
8
10.0000-99.999
1.5/0.15
9
100.000-999.99
2.5/0.2
1.000-99.9
10-3
5/1
Ls. H;
Lp. H;
tgδL (QR)
0.999 kHz. 0.100 kHz -0.999 kHz
1
000.01-099.99
10-12
1/0.1
2
100.000-999.99
0.5/0.05
3
1.00000-9.9999
10-9
0.1/0.01
4
10.00000-99.9999
5
100.0000-999.99
6
1.00000-9.9999
10-6
7
10.0000-99.999
0.2/0.02
8
100.000-999.99
0.5/0.05
9
1.00000-9.9999
10-3
1/0.1
10.000-9999
2/0.1
Ls. H;
Lp. H;
tgδL (QR)
1.00 kHz. 1.00 kHz -9.99 kHz
1
00.001-09.999
10-6
0.5/0.01
2
10.00000-99.9999
0.5/0.05
3
100.0000-999.99
0.2/0.02
4
1.00000-9.9999
10-3
0.02/0.002
5
10.0000-99.999
6
100.000-999.99
7
1.00000-9.9999
100
0.05/0.005
8
10.0000-99.999
0.1/0.01
9
100.000-999.99
0.25/0.025
1.00000-999.9
103
1/0.1
Ls. H;
Lp. H;
tgδL (QR)
10.0 kHz. 10.0 kHz -100 kHz
1
0.0001-0.9999
10-6
1.5/0.01
2
1.00000-9.9999
1/0.1
3
10.0000-99.999
0.5/0.05
4
100.000-999.99
5
1.00000-9.9999
10-3
0.2/0.02
6
10.0000-99.999
7
100.000-999.99
0.5/0.05
8
1.00000-9.9999
100
9
10.0000-99.999
100.000-999.99
1/0.1
1.000-99.9
103
5/1
Rs. Rр. tgφC. tgφL
1.00 kHz. 0.100 kHz -100 kHz
1
000.01-099.99
10-3
0.5/0.01
2
100.000-999.99
0.2/0.02
3
1.00000-9.9999
100
0.1/0.01
4
10.0000-99.999
0.05/0.005
5
100.000-999.99
0.02/0.002
6
1.00000-9.9999
103
7
10.0000-99.999
8
100.000-999.99
0.05/0.005
9
1.00000-9.9999
106
0.2/0.02
10.000-9999
0.5/0.1
 
   
Note:
- the initial measured values of AC bridge
R5083 AC are shown without regard to the initial parameters of the bridge and connecting cable for measurement object.
-
ƒon-the mainnormalfrequency;
-ƒn-additionalnormalfrequency;
AC bridge R5083 frequency range - from 0.1 kHz to 100 kHz;
Measurement time of the bridge when the main or additional measured value is changed within the range of measurements at frequencies of 1 kHz and above - not more than 0.7 s. at frequency from 0.1 kHz to 1 kHz - not more than 7 s;
Measurement time when the main unit of the measured value is changed for a less value at frequency from 1 kHz and above – not more than 70ms;
Time to select the measuring range of the bridge at frequency of 1 kHz and above - not more than 0.7 s. at frequency from 0.1 kHz to 1 kHz - not more than 4 sec.;
Operation mode setup time of AC bridge R5083 in normal operating conditions is not more than 1 hour after its turn on;
    The duration of continuous work of the AC bridge - 9 hours. Time for re-enable is not less than 1hour;
     The power consumed by AC bridge P5083 AC is not more than 100 V ∙A;
     AC Bridge is restored. and repairable device;
     The average lifetime service of  the AC bridge is at least 10 years;
     Considered a limiting condition in which the total repair costs exceed 60% of the cost of the bridge;
    AC bridge
R5083 normal conditions of use:
      - Temperature - 20 º C ± 1 º C;
      - Relative humidity - 30% -80%;
      - Atmospheric pressure - 84kPa-106. 7kPa;
      - Power supply frequency - 50 Hz ± 1 Hz;
      - AC voltage - 220V± 22V or 110V ± 11V;
     Dimensions of AC bridge
R5083 - not more than 440x158x405mm;
     Weight - not more than 16kg.

Photos P5083 (Р5083)

R5083 (P5083) device image.
R5083 (P5083) device image.
R5083 (P5083) front view.
R5083 (P5083) front view.
R5083 (P5083) side view.
R5083 (P5083) side view.
R5083 (P5083) rear view.
R5083 (P5083) rear view.
R5083 (P5083) overhead view.
R5083 (P5083) overhead view.
R5083 (P5083) bottom view.
R5083 (P5083) bottom view.


Яндекс цитирования