Bridge AC R5058 (R-5058, R 5058)
 
Bridge AC R5058 is designed to measure the capacitance and dissipation factor of capacitors and other objects of measurement, as well as for measuring conductivity and the residual capacity of high-resistance resistors.
 
Technical characteristics of equipment bridges AC R5058:
 
Operating frequency sub-ranges of measurement and the instrument allowed the measurement errors are shown in Table 1;
 
Table 1 Operating frequency sub-ranges of measurements and measurement errors allowed the device ac bridge R5058
 
Measured my option
Frequency
Sub-band measurements
Max. the value of the parameter measured emogo
Acceptable basic error limit device R5058 ac bridge
δo
Δ
Since, tgδ
0.05 kHz
10.00 nF-99, 99nF
0.5 kHz
1%
0,05 tgδ +5 ∙ 10 -3
100.0 nF-999, 9nF
1.000 uF-9 999mkF
10.00 mF-99, 99mkF
-
100.0 uF-999, 9mkF
2%
1kHz
0.02 pF, 99, 99pF
1kHz
0.5 0.004 (1000/S-1)%
-
100.0 pF-999, 9pF
0.5
1.000 nF-9, 999nF
0,02 tgδ +2 ∙ 10 -3
10.00 nF-99, 99nF
0.2
0,01 tgδ +1 ∙ 10 -3
100.0 nF-999, 9nF
0,005 tgδ +5 ∙ 10 -4
1.000 uF-9 999mkF
0.5
0,01 tgδ +1 ∙ 10 -3
10.00 mF-99, 99mkF
1%
0,02 tgδ +5 ∙ 10 -3
10kHz
0.02 pF, 99, 99pF
0.5 kHz
A 0.002 (1000/S-1)%
-
100.0 pF-999, 9pF
1%
0,02 tgδ +2 ∙ 10 -3
1.000 nF-9, 999nF
0.5
0,01 tgδ +2 ∙ 10 -3
10.00 nF-99, 99nF
100.0 nF-999, 9nF
1%
0,02 tgδ +5 ∙ 10 -3
C, go playlist
0.5 (tgδ ≤ 0,5)
10.00 nF-99, 99nF
31.4 mS
2
0.02 0.5 mS go playlist
100.0 nF-999, 9nF
0.314 mS
0.2 0.005 mS go playlist
1.000 uF-9 999mkF
3.14 mS
0.2 0.5 mS go playlist
10.00 mF-99, 99mkF
6.2 mS
-
1 (tgδ ≤ 1)
0.02 pF, 99, 99pF
628nSm
0.5 0.004 (1000/S-1)%
-
100.0 pF-999, 9pF
6.28 mS
0.5
0.05 0.005 mS go playlist
1.000 nF-9, 999nF
62.8 mS
0.2
0.02 0.05 mS go playlist
10.00 nF-99, 99nF
628mkSm
0.2
0.01 0.2 mS go playlist
100.0 nF-999, 9nF
6.28 mS
0.5
0.02 0.005 mS go playlist
1.000 uF-9 999mkF
62.8 mS
1%
-
10 (at tgδ ≤ 0,1)
100.0 pF-999, 9pF
6.28 mS
2%
0.02 0.1 mS go playlist
1.000 nF-9, 999nF
62.8 mS
0.02 mS 2 go playlist
10.00 nF-99, 99nF
628mkSm
0.02 mS go playlist 10
100.0 nF-999, 9nF
6.28 mS
-
L, Ro
0.5 (tgδ ≤ 0,5)
10.00 TR-99, 99Gn
15 kOhm
1.000 TR-9, 999Gn
1.5 kOhm
1 (tgδ ≤ 0,5)
1.000 TR-9, 999Gn
31.4 k
0.5
10.00 mH-999, 9mGn
3.14 k
0.2
Ro 0.02 0.005 ohms
1 (tgδ ≤ 1)
10.00 TR-99, 99Gn
628.0 Ohm
Ro 0.02 0.5 ohms
1.000 TR-9, 999Gn
62.80 ohms
Ro 0.05 0.05 Ohm
100.0 uH-999, 9mkGn
6.280 ohms
0.5
Ro 0.05 0.20 Ohm
0.03 mH, 99, 99mkGn
628.30 milliohms
0,5 +0,003 (1000/L-1%)
-
10 (at tgδ ≤ 0,1)
100.0 mH-999, 9mGn
9.99 k
1%
Ro 0.05 0.05 Ohm
10.00 mH-99 99mGn
999 Ohm
0.05 Ro 5 ohms
1.000 mH-9, 999mH
99.9 Ohm
Ro 0.05 0.5 ohms
100.0 uH-999, 9mkGn
9.99 Ohm
-
R, LO
1 (for τ ≤ 100mks)
1.000 MW-9, 999 MW
-
-
100.0 k-999, 9 k
1.00 Mr
1%
10.00 k-99, 99 ohm
0.5
LO 0.05 0.005 TR
1.000 ohm-9, 999 ohms
999.9 mH
0.2
LO 0.02 0.5 mH
100.0 ohm-999 ohm 9
99.99 mH
0.05 LO 0.05 mH
10.00 ohm-99, 99 ohm
9.999 mH
0,2 +0,05 (100/R-1)%
-
1.000 ohm-9, 999 ohms
999.9 uH
0,5 +0,2 (10/R-1)%
10.0 ohm-999 ohm 9
99.99 mH
10 (for τ ≤ 5mks)
10.00 k-99, 99 ohm
0.10 Mr
2%
1.000 ohm-9, 999 ohms
99.9 mH
LO 0.05 0.5 mH
100.0 ohm-999 ohm 9
9.99 mH
1%
0.05 LO 0.05 mH
R, τ L
A
10.00 k-99, 99 ohm
9.99 ms
-
1.000 ohm-9, 999 ohms
0.5
100.0 ohm-999 ohm 9
0.2
0,02 τ L +0,5 ms
10.00 ohm-99, 99 ohm
0,2 +0,1 (100/R-1)%
-
R, τ with
A
1.000 MW-9, 999 MW
99.99 ms
1%
100.0 k-999, 9 k
0.5
0,02 τ of 0.5 ms
10.00 k-99, 99 ohm
0.2
0,01 τ of 0.1 ms
1.000 ohm-9, 999 ohms
100.0 ohm-999 ohm 9
0.5
0,01 τ of 0.2 ms
10.00 ohm-99, 99 ohm
0,02 τ of 0.5 ms
10
10.00 k-99, 99 ohm
1.99 ms
2%
-
1.000 ohm-9, 999 ohms
1%
0,02 τ of 0.5 ms
100.0 ohm-999 ohm 9
-
G, Co
0.05 (at tgδ ≤ 0,5)
100.0 mS-999, 9mkSm
99.99 nF
1.000 mS-9 999mSm
999.9 nF
10.00 mS-99 99mSm
9.999 uF
1 (tgδ ≤ 1)
1.000 mS-9 999mkSm
999.9 pF
0.02 With 0.2 pF
10.00 mS-99 99mkSm
9.999 nF
0.5
0.01 Co 0.02 nF
100.0 mS-999, 9mkSm
99.99 nF
0.2
1.000 mS-9 999mSm
999.9 nF
0.5
0.01 Over 0.2 nF
100.0 mS-999, 9mSm
99.99 mF
2%
-
10 (at tgδ ≤ 0,1)
100.0 mS-999, 9mkSm
9.99 nF
1.000 mS-9 999mSm
99.9 nF
1%
10.00 mS-99 99mSm
.999 UF
C
A
± (0,05 ÷ 30)
A
A
(0.02 P 0.05)%
 
Time measurements at frequencies 10kHz, 1kHz and 50Hz, respectively less than 0.1 s, 0.2 s and 5 s;
The choice of sub-band measurement device ac bridge R5058 - automatic;
Selecting the operating mode and operating frequency - Manual and remote;
The device is a bridge AC R5058 has concluded the results of measurements in parallel BCD 8-4-2-1 with signal levels corresponding to:
- 12V ± 2,4 V - for a logical "1";
- 0 V ± 2,4 V - for a logical "0";
The bridge has three types of launch:
manual
- Automatic with an adjustable loop in the range of 0.5-10s;
- External, controlled by a series of pulses of amplitude 2.5 V-6 V;
Changing the limit permitted by the basic instrument error R5058 ac bridge is less than half the values ​​shown in Table 1 as:
- Changes in ambient temperature relative to the nominal value of 20 º C ± 5 º C in the range of 10 º C to 35 º C for every 10 K change in temperature;
- Changing the voltage from +10% to -15% of the nominal value of 220 ± 2%;
- The presence of an external magnetic field of frequency 50 Hz ± 1 Hz with intensities up to 80 A / m;
Operation mode setup time unit bridge AC R5058 - not more than 1 minute after turning it on;
Continuous operation time - up to 8 hours;
The manufacturer guarantees the quality of the device ac bridge R5058 technical requirements specifications, subject to the consumer the terms and conditions of storage, transportation, installation, operation, established specifications and operational documentation.
 