Electrical measuring instruments >> measuring bridges

R5016 Bridge AC R5016

R5016

69 999 RUB
28 000 UAH
available:
1 piece

Online prices include VAT.

For order is possible effectively delivery of any quantity.

Bridge AC R5016 (R-5016, R 5016, p5016, p-5016, p 5016)

For measuring the capacitance C and the loss tangent tg δ in parallel substitution inductance L and the loss tangent tg δ in series substitutional resistance R and the tangent of the phase shift tg φ.
Measurement ranges main parameters: C - from 10-14 F to 10-4 L - 10 -7 H to 10 2 GN; R - 10 -2 to 106 ohms.
Measuring ranges of related parameters: tgδ - from 10-4 to 1 and from 10-3 to 1; tgφ - from-3 to +10 ± 1.
Dimensions 490x210x495 mm, weight 35 kg.

Bridge AC R5016 designed to measure capacitance C and the loss tangent tg δ in parallel displacement, the Land inductance dissipation factor tg δ of substitution in series, resistance R and the tangent of the phase shift tg φ.
In terms of mechanical and climatic influences Bridge AC R5016 correspond to GOST 22261-76 (group 2).
Main technical characteristics
Measurement ranges main parameters: C - from 10 to 14 F to 10 F -4; L - 10 -7 H to 10 2 GN; R - 10 -2 to 10 ohms 6 ohms.
Measuring ranges of related parameters: tgδ - from 10 -4 to 1 and from 10 -3 to 1; tgφ - from -3 to +10 ± 1.
The number of decimal places of reference for C, L, R-five and tgδ and tgφ - four.
Bridge AC R5016 has three measurement modes: search mode, a mode of repeated measurements and tracking mode.
Starting the bridge manual, automatic and external. Selection of sub-band measurements automatically.
The device provides output measurements in BCD parallel-serial code 8-4-2-1.
Logic "1" corresponds to the voltage from 2.4 to 5.25 V, the logical "0" - from 0 to 0.4 V.
Pulse parameters, "Full-scale": the amplitude of 2.4 to 5.25, the duration from 4 to 10 ms, the rise time is not more than 0.5 microseconds.
Time measurement is not more than 1.5 sec. Warm-up time of 1 min.
The limit of additional error appliance Bridge AC R5016 caused a deviation from the ambient temperature (20 ± 1) ° C to any temperature within the operating temperature for each 10 K temperature deviation is equal to the value of 0,5 δ 0 for δ 0 ≥ 0,1 %, δ 0 - at 0,05% ≤ δ 0 <0.1%, and 0 at 2δ 0,02% ≤ δ 0 <0.05%.
The limit of additional error under the influence of an external magnetic field, 50 Hz, 80 A / m, as well as the bridge supply voltage deviates by ± 10% of nominal value equal to the value 0,5 δ 0.
Power supply Bridge AC R5016 is powered by the AC, 50 Hz ± 1% 220V ± 10% THD up to 5%.
Power consumption is less than 200 VA.
Dimensions 490h210h495 mm, weight 35 kg.
Mean time between failure is not less than 1250 hours
 
Measured parameter (main) Measurement range of the main parameters on the frequency, kHz, equal
1kHz 5 kHz and 10 kHz 50kHz
Capacitance, F 0-10 -9 F 0-10 -10 F 0-10 -11 F
10 F-10 -9 -8 F F-10 -10 10 -9 F F-10 -11 10 -10 F
10 -8 -7 F-10 F 10 F-10 -9 -8 F F-10 -10 10 -9 F
F-10 -7 10 -6 F 10 -8 -7 F-10 F 10 F-10 -9 -8 F
10 -6 -5 F-10 F F-10 -7 10 -6 F 10 -8 -7 F-10 F
F-10 -5 10 -4 F 10 -6 -5 F-10 F F-10 -7 10 -6 F
Inductance, H 10Gn-100Gn 1Gn-10Gn 0.1 GN-1Gn
1Gn-10Gn 0.1 GN-1Gn 0.01 Mr-0, 1Gn
0.1 GN-1Gn 0.01 Mr-0, 1Gn 10 -3 10 -2 GN-GN
0.01 Mr-0, 1Gn 10 -3 10 -2 GN-GN 10 -4 to 10 -3 Mr Mr
10 -3 10 -2 GN-GN 10 -4 to 10 -3 Mr Mr 10 -5 to 10 -4 GN-GN
0-10 -3 Mr 0-10 -4 Mr 0-10 -5 Mr
Resistance in ohms 105 Ohms 10 6 ohms 105 Ohms 10 6 ohms 105 Ohms 10 6 ohms
104Ohm-105Ohm 104Ohm-105Ohm 104Ohm-105Ohm
103Ohm-104Ohm 103Ohm-104Ohm 103Ohm-104Ohm
102Ohm-103Ohm 102Ohm-103Ohm 102Ohm-103Ohm
-100 Ohms 10 Ohms 10 Ohm -100 Ohm -100 Ohms 10 Ohms
0-10 ohms 0-10 ohms 0-10 ohms
Measured parameter (main) Maximum permissible error of measurement δ 0 for tgδ ≤ 0,05 and tgφ ≤ 0,05
Relative primary parameter δ 0,%, at a frequency kHz, equal
1kHz 5 kHz 10kHz 50kHz
Capacitance, F 0,1 +100 / α 0,2 +1000 / α 0,5 +10000 / α 5 +10000 / α
0,02 +100 / α 0,1 +1000 / α 0.2 F 2F
0.02 F 0.05 F 0.1 F 2F
0.05 F 0.1 F 0.2 F 5F
0.1 F 0.2 F 1F -
0.2 F - - -
Inductance, H - - - -
0.5 Mr 2Gn 5Gn -
0.2 Mr 1Gn 2Gn -
0.05 Mr 0.2 Mr 0.5 Mr 5Gn
0.1 Mr 0.5 Mr 1Gn 5Gn
0,1 +500 / α - - -
Resistance in ohms 0.5 Ohm 2 ohms - -
0.2 Ohm 1 ohm 2 ohms 5 ohms
0.05 Ohm 0.2 Ohm 0.5 Ohm 2 ohms
0.1 Ohm 0.5 Ohm 1 ohm 2 ohms
0,2 +5000 / α 0,5 +10000 / α 1 ohm 5 ohms
50000 / α - - -
Measured parameter (main) Maximum permissible error of measurement δ 0 for tgδ ≤ 0,05 and tgφ ≤ 0,05
absolute associated parameter Δ o at frequency kHz equal
1kHz 5 kHz 10kHz
Capacitance, F - - -
± (0,01 tgδ + 2.10 -4) ± (0,02 tgδ + 5.10 -4) ± (0,05 tgδ + 1.10 -3)
- - -
- - -
Inductance, H - - -
- - -
± (0,02 tgδ + 5.10 -6) ± (0,05 tgδ + 1.10 -3) ± (0,1 tgδ + 2.10 -3)
- - -
Resistance in ohms - - -
± (0,02 tgφ · 10 -3) ± (0,02 tgφ + 1.10 -3) ± (0,02 tgφ + 1.10 -3)
± (0,01 tgφ + 5.10 -4)
- - -
- - -

Note. 1. When tgδ> 0,05 or tgφ> 0,05 C measurement error, the Land Kuvelichivaetsya to 0,1 tgδ f or 0,1 tgφ f, where f - frequency, kHz.
2. α - the reading of the bridge without a comma.
3. A dash means that the error is not standardized.
4. The errors do not include the initial parameters of the bridge.