Electrical measuring instruments >> measures and boxes >> inductance

R5009 (P5009)

39 999 RUB
14 000 UAH
available:
6 pieces

Online prices include VAT.

For order is possible effectively delivery of any quantity.

Measure the mutual inductance R5009

Also this device can be called: R-5009, R 5009, P5009, P-5009, P 5009.

 

R5009 measure mutual inductance is used as a model to measure the inductance of AC circuits at work in the laboratory calibration of industrial enterprises and research institutions.

 

Measures of the mutual inductance of the two windings consists arranged concentrically in a plastic housing.

 

Specifications R5009:

 

Rated inductance - from 100 mcH to 10 mH.

Rated frequency - 5 kHz.

Accuracy class - 0,1.

 

Type Rated inductance Value of the parameters of the primary winding Maximum resistance of the secondary winding
Nomination flax current Maximum voltage Maximum resistance
R5009/1 100 mcH 2,5 A 100 V 1,0 Ohm 1,0 Ohm
R5009/2 300 mcH 2,0 A 100 V 1,5 Ohm 1,5 Ohm
R5009/3 500 mcH 2,0 A 100 V 2,0 Ohm 2,0 Ohm
R5009/4 1 mH 1,5 A 300 V 2,5 Ohm 2,5 Ohm
R5009/5 3 mH 1,0 A 300 V 5,0 Ohm 5,0 Ohm
R5009/6 5 mH 0,5 A 100 V 10 Ohm 16 Ohm
R5009/7 10 mH 0,5 A 300 V 15 Ohm 25 Ohm

 

Certification measures carried out at a frequency of R5009:

- R5009/1, R5009/2 - 5 kHz; 30 kHz; 50 kHz;

- R5009/3, R5009/4, R5009/5 - 5 kHz; 10 kHz; 30 kHz;

- R5009/6 R5009/7- 1 kHz; 5 kHz; 10 kHz.

 

Maximum permissible error R5009 measures, expressed as a percentage of the nominal value under normal operating conditions and frequency deviation from nominal by no more than ±0,1% - less than ±0,1%.

 

Accuracy evaluation measures, expressed as a percentage of the nominal value of the mutual inductance:

- at 1 kHz and 10 kHz - not more than ±0,1%;

- at 30 kHz and 50 kHz - not more than ±0,2%.

 

Phase error at nominal frequency - no more than 3·10-3 rad.

Changing the temperature coefficient measures R5009 temperature deviates from the normal - no more than 0,005% at 1 K.

 

Test voltage - 2 kV.

MTBF - not less than 49 500 hours.

Average life - at least 6 years.

Dimensions - 200×210×100 mm.

Weight measures R5009 - not more than 1,6 kg.

 

Operating conditions R5009:

 

Ambient temperature - +10° C to +35° C.

Relative humidity - 80%.

Photos: R5009 (P5009)

R5009 (P5009) measure image.
R5009 (P5009) measure image.
R5009 (P5009) front view.
R5009 (P5009) front view.
R5009 (P5009) side view.
R5009 (P5009) side view.
R5009 (P5009) rear view.
R5009 (P5009) rear view.
R5009 (P5009) overhead view.
R5009 (P5009) overhead view.
R5009 (P5009) bottom view.
R5009 (P5009) bottom view.


Яндекс цитирования