Radiometric devices >> lines measuring

R1-22

available:
1 piece
For order is possible effectively delivery of any quantity.

Line measurement R1-22 (R122, R1 22)
P rednaznacheny to measure the standing waves in coaxial paths cross section of 16/6, 95 mm.
1-7,5 GHz frequency range.
Net KstU 1.02 (up to 2 GHz) 1.025 (5.5 GHz), 1.04 (7.5 GHz).
Variability communication with the field lines probe: 1.2% (5.5 GHz), 2% (7.5 GHz).
RF path 50 ohm (16/7 mm)

 

Line measurement R1-22 is designed to measure the standing wave paths in coaxial cross-section 16/6, 95 mm.

 

Line R1-22 allows you to measure the characteristics of loads with high-frequency connectors of the "plug" and "socket".
Line measurement R1-22 stores the specification in the temperature range from 278 to 313 K (5 to 40 ° C) and humidity of 95% at a temperature of 303 K (30 ° C).
R1-22 belong to the class 2 GOST 11294-74.
Technical data
Operating frequency range from 1.0 to 7.5 GHz. Characteristic impedance 50 ohms line, microwave connectors lines correspond to the type II (on the one hand, "fork", the other "socket") GOST 13317-73.
Own standing wave ratio line (K st U) is not more than:
1.02 in the range from 1 to 2 GHz in the range of 1.025 to 5 over 2 GHz in the range of 1.04 to 7.5 than 5.0 GHz.
Variability communication with the field lines probe max: 1.2% in the range from 1.0 to 5.5 GHz, 2% over the range 5.5 to 7.5 GHz. Shunt Relative conductivity probe is not more than 0,013. Repeat accuracy of the probe is not more than 0.04 mm.
Margin of the probe along the measuring line measuring R1-22 is not less than 225 mm. A) For short-Art U loads of at least 50 at a frequency of 1 GHz,
30 at a frequency of 7.5 GHz.
b) impedance should be 50 ohms ± 0,7%.
c) measuring the position of the plunger error not exceeding 0.04 mm.
a) To Hg U mismatched loads 1,2 ± 0,1 in the frequency range from 2.0 to 7.5 GHz.
b) Impedance 50 ohm ± 0,7%.
Guaranteed number of strokes mechanisms strokes at least 30,000.
Guaranteed number of moves and short-circuiting of the absorbing elements load of at least 10,000.
Normal operating conditions: temperature 293 ± 5 K (20 ± 5 ° C), relative humidity of 65 ± 15%, atmospheric pressure 100 ± 4kN / m 2 (750 ± 30 mm Hg. Cent.).
Operating conditions: temperature from 278 to 313 K (5 to 40 ° C), relative humidity 95% at temperatures up to 303 K (30 ° C), atmospheric pressure 100 ± 4 kN / m 2 (750 ± 30 mm Hg. cent.).
Overall dimensions of the line R1-22
440h300h233 mm. Dimensions of packing box line
512h455h344 mm. Dimensions of packing box loads
354h68h214 mm. Overall dimensions of a short-circuited load number 1
Ø30x 293 mm. Overall dimensions of a short-circuited load number 2 Ø32h305 mm.
Dimensions mismatched load number 3
Ø32x 297 mm. Dimensions mismatched load number 4
Ø 30h293 mm. Overall dimensions of shipping containers 740h606h526 mm.
Mass lines R1-22 is not more than 11 kg. Mass line in the packing box not more than 35 kg. The weight of one load of 0.8 kg. Weight loads in the packing box is not more than 5 kg. Line mass in transport containers not exceeding 65 kg.
The average number of strokes uptime mechanisms meter run at least 45,000, short-circuiting and absorbing elements of loads of at least 10,000.