Electrical measuring instruments >> measures and boxes >> other measures and stores

N4-100 (H4-100)

80 353 RUB
21 016 UAH

Online prices include VAT.

For order is possible effectively delivery of any quantity.

Measure the voltage N4-100

Also, this device can be called: N4100, N4 100, N4-100, h4100, h4 100.

N4-100 measure voltage - DC voltage source of high-precision values ​​of 1 V, 1,0184 V, 10 V is intended for research, inspection and calibration of equipment.

At the heart of the instrument lies robots reference voltage source, which is described as a chip comprising a zener reference resistive heater and temperature sensing transistor.

Scope: electronics, automation, instrumentation.

Specifications N4-100:

The actual value of the DC voltage measures:

- 1 V - 1±0,0002 V;

- 1,0184 V - 1,0184±0,0002 V;

- 10 V - 10±0,0002 B.

Instability (change) 1,0184 V voltage:

- 4 hours - 1 mV ;

- Within 30 days - 2 mV;

- 90 days - 5 mV;

- 1 year - 10 mV.

Instability (change) voltage 1:

- 4 hours - 1 mV ;

- Within 30 days - 2 mV;

- 90 days - 5 mV;

- 1 year - 10 mV.

Instability (change) 10 V:

- 4 h at - 10 mV;

- Within 30 days - 20 mV;

- 90 days - 30 mV;

- 1 year - 40 mV.

Voltage change N4-100 after shutdown:

- 1 V - ± 1 mV;

- 1,0184 V - ± 1 mV;

- 10 V - ± 10 mV.

Changing the output voltage from the temperature change:

- 1 V, 1,0184 V - ± 3 mV;

- 10 V - ± 20 mV.

RMS output voltage of the variable components of N4-100 in the frequency range from 0,01 kHz to 10 kHz:

- Output 1 V - more than 20 mkV;

- In the output 1,0184 V - no more than 20 mkV;

- Output 10 V - more than 40 mkV.

Output impedance of the device measure the voltage is:

- Output 1 V - less than 1 Ohm;

- Output 1,0184 V - max 1 mW;

- Outlet 10 V - less than 1 kOhm.

Output current N4-100 outlet 10 V - not less than 12 mA.

Operating mode - no less than 24 hours.

Dimensions - 105x285x360 mm.

Weight:

- Without packaging - no more than 6 kg;

- Package - 10 kg.

Operating conditions N4-100:

Ambient temperature

- Operating - from 10º C to 35º C;

- Norm - 21 ± 2º C.

Relative humidity at 25º C:

- Working - up to 80%;

- Rate - from 30% to 80%.

Atmospheric pressure - from 106,7 kPa to 84 kPa (630 mm hg to 800 mm hg).

AC voltage - 230±23 V.

The frequency of the AC - 50±0,4 Hz.

Power consumption measures voltage - 20 V·A.

Antidecoying N4-100 - Class B.

Effect of an external DC or AC magnetic field frequency of 50Hz - not more than 40 A/m

Hours - 60 hours.

MTBF - not less than 12,000 hours.

Average yield N4-100 - at least 10,000 hours.

Average time disaster recovery - no more than 3 hours.

Photos: N4-100 (H4-100)

N4-100 (H4-100) device image.
N4-100 (H4-100) device image.


Яндекс цитирования