Absolute Pressure Gauge MPAK-15 (MPAK15, MPAK 15)
 
Absolute Pressure Gauge MPAK-15 is designed for checking and calibration of the absolute and gauge pressure, barometers and vacuum gauges.
Scope - Calibration gauge and absolute pressure, barometers and vacuum gauges in metrological centers and laboratories of the TRC.
 
Technical characteristics of the instrument gauges for absolute pressure MPAK-15:
 
Measurement range - 0.133 kPa, 400kPa (1mm.rt.st.-3000mm.rt.st, from 0,133 ∙ 10 -2 kgf / cm 2 to 4kgs/sm 2);
Acceptable basic error limit device MPAK-15 gauge absolute pressure:
- In the range from 0.133 kPa to 13.3 kPa (1mm.rt.st. to 100mm.rt.st., From 0,133 ∙ 10 -2 kgf / cm 2 to 13,3 ∙ 10 -2 kgf / cm 2) - ± 6,65 ∙ 10 -3 kPa (± 0,05 mm Hg; ± 6,65 ∙ 10 -5 kg / cm 2);
- In the range from 13.3 kPa to 133kPa (from 100mm.rt.st. To 1000mm.rt.st., From 13,3 ∙ 10 -2 kgf / cm 2 to 133 ∙ 10 -2 kgf / cm 2) - ± 13,3 ∙ 10 -3 kPa (± 0,1 mm Hg; ± 13,3 ∙ 10 -5 kg / cm 2);
- In the range from 133kPa to 400kPa (from 1000mm.rt.st. To 3000mm.rt.st., From 1.3 kg / cm 2 to 4kgs/sm 2) - ± 0,01% of the actual value of the measured pressure;
The nominal value of the difference between the above areas of measurement and siloperedayuschego (compensating) piston unit of absolute pressure gauge MPAK-15 - 2 cm 2;
The threshold of response - 3.3 Pa (0.025 mm Hg; 0,336 ∙ 10 -4 kgf / cm 2);
The working fluid - oil instrumentation IMP;
Working fluid absolute pressure gauge device MPAK-15 on one filling - about 50 cm 3;
Power supply is AC 220V ± 22 V and a frequency of 50 Hz ± 0,4 Hz;
The maximum power consumption - 50W;
Air is supplied to the pressure gauge must comply with pollution 0;
The device is an absolute pressure gauge MPAK-15 must be installed on a solid table, which excludes the effect of vibration on the operation of the device;
The table should be set no closer than two meters from the windows, heating units and other sources of heat or cold;
Dimensions of the unit of absolute pressure gauge MPAK-15 - 470h300h360mm;
Weight - no more than 25kg;
The manufacturer guarantees the quality of the instrument gauge absolute pressure MPAK-15 technical requirements specifications, subject to the consumer the terms and conditions of storage, transportation, installation, operation, established specifications and operational documentation.