Laboratory equipment >> freight piston pressure manometers


134 461 RUB
33 616 UAH

Online prices include VAT.

For order is possible effectively delivery of any quantity.
    Manometer gas deadweight MGP-10 
    Also, this device can be called: MGP10, MGP 10.

    Gas gauges Deadweight MGP-10 - designed to create and accurately measure pressure and vacuum gas.
    At the heart of the instrument robots MGP-10 is the principle of "non-compacted piston" - the total weight of the piston weighbridge, and installed on the balances of goods measured by the pressure acting on the lower end of the piston.
    Scope of the MGP-10: a standard for verification and calibration of measuring the excess pressure gauges, transducers, model and technical gauges and other measuring instruments (except oxygen SI) under laboratory conditions.


    Accuracy class MGP-10 - 0.02 0.05 (Table 1).
    Upper limit of measurement (playing) - 1 MPa (10 kgf / cm 2).
    The lower limit of measurement (playback) - 0.02 MPa (0.2 kgf / cm 2).
    Readability create pressure, provides a standard set of goods - 5 kPa.
    The piston stroke length - at least 10 mm.
    Nominal area of the piston - 1 cm 2.
    Maximum deviation from the nominal value of the reduced area of the piston - 1%.
    Working fluid - air.
    Dimensions - 500h400h240 mm.
    Weight without cargo - not more than 25 kg. 
Table 1
Appointment setting MGP-10 cells. ie 0.02 MGP-10 cells. ie 0.05
Maximum permissible relative error 0.02% 0.05%
Sensitivity threshold MGP-10 not more than 2 Pa not more than 5 Pa.
Lowering speed of the piston less than 1 mm/min not more than 2 mm/min
Free rotation of the piston gauge MGP-10 not less than 5 min at least 4 minutes
    Terms of use of the appliance gas gauge deadweight MGP-10:

    Ambient - 15 º C to 30 º C.
    Relative humidity - 80%.
    Atmospheric pressure - from 84 kPa to 106.7 kPa.
    AC voltage - 220 V.
    The frequency of the AC - 50 Hz.
    MTBF - 3000 hrs
    Average life - 8 years.
    The manufacturer provides a guarantee the compliance of the gas gauge deadweight MGP-10 all the technical requirements provided the customer terms and conditions of use, storage and transportation, installed documentation manuals.

Яндекс цитирования