Laboratory equipment >> devices for thermophysical and thermal measurements

ITP-MG4 «100» ITP-meter heat MG4

ITP-MG4 «100»

For order is possible effectively delivery of any quantity.
ITP-MG4 "100" meter heat  (ITPMG4, ITP MG4, ITP-MG4)
 
ITP-MG4 "100" meter heat  is used to determine the thermal conductivity and thermal resistance of building materials and materials for thermal insulation of industrial equipment and pipelines at steady state and the method of thermal probe.
 
Technical characteristics of the devices Thermal Conductivity ITP-MG4 "100":
 
Measurement range of the thermal conductivity in stationary mode - 0.02 W / m · K-1, 5 W / mK;
Measuring range of the instrument measuring the thermal resistance of the heat-ITP MG4 "100" - 0.01 m 2 K/Vt-1 5 m 2 · K / W;
Measurement range of the thermal conductivity by heat probe - 0.03 W / m · K-1 0W / m K;
Limit of basic relative error of determining the coefficient of thermal conductivity and thermal resistance of the device under steady state thermal conductivity meter ITP-MG4 «100» - ± 5%;
Limit of basic relative error in determining the thermal conductivity of the thermal probe method - ± 7%;
Dimensions of the test specimen - 100h100h (3-28) mm;
Memory capacity measurements instrument measuring conductivity ITP-MG4 "100":
- Steady state - 200;
- The method of thermal probe - 200;
Measurement time measuring device heat ITP-MG4 "100":
- At steady state - no more than 60 minutes;
- By heat probe - no more than 10 minutes;
Dimensions of the device measuring the thermal conductivity of ITP-MG4 "100":
- Electronic block - 175h90h30mm;
- Heating system - 155h200h210mm;
Weight of the device measuring the thermal conductivity of ITP-MG4 "100" - no more than 4 kg;
The device allows to determine the thermal conductivity and thermal resistance of the material at an average temperature of the sample from +15 ° C to +42,5 ° C, providing automatic temperature cooler and heater and temperature control during testing;
Obtained during the measurement of thermal conductivity measuring device ITP-MG4 "100" information is automatically backed up and marked with the date and time of measurement;
The devices can be equipped with a thermal probe for operational control of the thermal conductivity of materials during their production and use, as well as the inspection of buildings and structures, have a connection to a PC;
The manufacturer guarantees that the quality of the instrument measuring the thermal conductivity of ITP-MG4 "100" to the technical requirements specifications provided the customer terms and conditions of storage, transportation, installation, operation, and its specified operational documentation.