Laboratory equipment >> dosimeters, radiometers

IMD-1 Meter dose IPD-1.

IMD-1

For order is possible effectively delivery of any quantity.
Meter dose IMD-1 (IMD1, IMD 1)
 
Meter dose IMD-1 is designed to measure the field scattered by daylight and darkness of the exposure vine gamma radiation in the energy range from 0.08 MeV to 3MeV and detect beta radiation.
 
Technical characteristics of the instrument gauges dose IMD-1:
 
The device has three variants, which are listed in Table.1;
 
Table 1: Three versions of the instrument measuring dose IMD-1
Embodiment of the
OKP code
Name of
Designation
IPD-1A
ZHSH1.289.199
43 6715 1032
IPD-1R
ZHSH1.289.199-01
43 6715 1033
IPD-1C
ZHSH1.289.199-02
43 6715 1034
 
Measurement range of the instrument measuring dose IMD-1 is:
- For the IPD-1A - 0.01 R / h to 999R / h;
- For IPD and IPD-1R-1C - 0.01 mR / h to 999R / h;
which is divided into two sub-bands:
- «MR / h» with outside dimensions of 0.01 mR / h to 999mR / h;
- «R / h» with outside dimensions of 0.01 R / h to 999R / h;
The basic error of the meter relative to the measured value does not exceed ± 25%, plus or minus one junior level when measured in the sub-bands of 0.1 mR / h to 999mR / h and 0.1 R / h to 999R / h;
Note. Subranges from 0.01 mR / hr to 0.1 mR / hr and 0.01 R / hr to 0.1 R / h is an indicator and the main error in these sub-bands the instrument measuring dose IMD-1 is not standardized;
Meter IPD-1 provides the audio alarm when the exposure dose rate 0.1 mR / h and 300mR / h on sub-band «mR / h» and 0.1 R / h and 300R / h - on sub-band «R / h»;
The instrument measuring dose IMD-1 preserves the performance (signal overflow detecting circuit de-excitation of the three nines in the senior ranks of digital display) when exposed to ionizing radiation exposure dose to the 100R / h - with a detection unit (IMD Meter for 1P-and IPD-1C) and to 1000R / h - without it;
Additional error due to the energy dependence of the sensitivity of the meter in the energy range from 0.08 MeV to 1.25 MeV does not exceed ± 25% relative to the mean value of sensitivity;
Additional error instrumentation gauges dose IMD-1 due to the anisotropy in the gamma-radiation energy 0.667 MeV does not exceed ± 25% for all of the provisions of the meter, but the direction of the radiation from the battery compartment - for remote measuring IPD-1-3 and the end surfaces - for detection unit IPD-1-1 (meters IPD IPD-1R and-1C);
Operation mode setup time meter - 1 minute;
The measurement time shall not exceed:
- For measuring IPD-1A:
- 15s - in the range of 0.01 R / h to 9.99 R / hr;
- 1.5 to - ranging from 10P / h to 999R / h;
- Meters for IPD and IPD-1R-1C:
- 60s - on the sub-band «mR / h» - 0.01 mR / hr to 9.99 mR / h;
- 6s - in sub-band «mR / h» - from 10mR / h to 999mR / h;
- 15s - on the sub-band «R / h» - 0.01 R / h to 9.99 R / hr;
- 1.5 to - on the sub-band «R / h» - from 10P / h to 999R / h;
Time of continuous operation of the instrument measuring dose IMD-1 from one set of elements - at least 100 hours.In this case, the additional error does not exceed ± 10%;
Feeding the meter is carried out:
- IPD-1A, IPD-1R - from four cells in series A-343 "Prima" with a nominal voltage of 6 V and on-board DC power or rechargeable batteries with a voltage of 10.8 V to 30 V;
- IPD-1 C - from four cells in series A-343 "Prima" with a nominal voltage of 6 V, on-board DC power or rechargeable batteries with a voltage of 10.8 V to 30V AC to 220V ± 22V with frequency 50 Hz or 400 Hz;
The instrument measuring dose IMD-1A, meter dose IMD-1R, meter dose IMD-1S has a device that signals cells to discharge about voltage 4V ± 0,1 in the inclusion on the panel indicator light;
Gauges and 1P-IPD IPD-1C provide the ability to detect beta radiation with an average energy;
Overall dimensions and weight of blocks included in the instrument meter dose IMD-1 are shown in Table 2;
 
Table 2: Size and weight of blocks included in the instrument meter dose IMD-1
Name of
Dimensions, not more
Weight, not more
Remote measuring
364h131h66mm
1.5 kg
The detection unit
60h171h70mm
0.6 kg
Power Supply IPD-1-2
150h110h55mm
0.7 kg
Power Supply IPD-1-6
143h135h73mm
1.1 kg
Bar
(305-720) h50mm
0.3 kg
Cord
1255h50h30mm
0.22 kg
Tourniquet
1500mm
0.14 kg
Tourniquet
20m
1.65 kg
Tourniquet
1535mm
0.13 kg
Tourniquet
1535mm
0.13 kg
Connecting cable
1535mm
0.15 kg
Box
389h388h160mm
2.8 kg
Box
564h388h160mm
3.1 kg
Box
564h454h160mm
5.5 kg
 
Meter provides a functional test measuring control - from the internal oscillator and built-in source of beta radiation detection unit (IMD Meter for 1P-and IPD-1C) - from background radiation;
The instrument measuring dose IMD-1 stably operates in the range of operating temperatures from -50 º C to +50 ° C and remains serviceable after exposure:
- Sinusoidal vibration at frequencies from 5 Hz to 80 Hz with an acceleration of 39m / s 2 (4g), and blocks of IPD-1 and IPD-2-1-6 (IMD Meter-1C) - from 5Hz to 300Hz with acceleration of 49m / s 2 (5g);
- Strikes at transportation in all modes of transport packaging for any distance;
- Low atmospheric pressure to 90 mm Hg. of Art. (At aviatransportirovanii) at -50 ° C;
- Changes in ambient temperature from -60 º C to +65 ° C;
- Solar radiation;
- Sand and dust;
- Fall from a height of 0.75 m on the ground;
- Working solutions for special handling (with the exception of solutions based on dihloretana).
The instrument measuring dose IMD-1A, meter dose IMD-1R, meter dose IMD-1S operates steadily at destabilizing factors with an additional error shown in Table 3;
 
Table 3
Destabilizing factor
Additional error of the instrument measuring dose IMD-1, not more
Temperature change in the temperature range for every 10 º C
± 5%
The relative humidity of 100% at ± 25 º C
± 10%
Change in atmospheric pressure from normal to 450mm Hg. of Art. (6 ∙ 10 4 Pa)
± 8%
The impact of frost and dew
± 25%
The impact of a constant magnetic field strength of up to 318A / m (4E)
5%
 
The manufacturer guarantees the quality of the instrument measuring dose IMD-1 technical requirements specifications subject to user terms and conditions of storage, transportation, installation, operation, established specifications and operational documentation.