High-frequency generator GK4-21A (GK421A, GK-4-21A, GK 4 21A, GK4 21A, GK-421A, GK 421A, GK421-A, GK421 A)

High-frequency generator GK4-21A is designed to test the parameters of wireless devices.
Technical characteristics of the instrument high-frequency generator GK4-21A:
The ambient air temperature - 30 º C. .. 50 ° C;
Relative humidity - 90% at 30 ° C;
Atmospheric pressure - 60kPa ... 106kPa (460mm.rt.st .... 800mm.rt.st.)
Voltage:
- 220V ± 22V with a frequency of 50 Hz ± 0,1 Hz;
- 115 V ± 5,75 V and 220 V ± 11V with a frequency of 400 Hz ± 12 Hz;
Frequency - 1765MGts-2564MGts;
Margin at the edges of the frequency range of high frequency signal generator unit GK4-21 - not less than 0.1%;
Error of frequency on the scale - less than ± 1%;
The instability of the oscillator frequency after 30 minutes for any samoprogreva 15 minutes - no more than ± 5 ∙ 10 -4 ƒ;
Additional time after warm-frequency tuning for this instability - not more than 5 minutes;
The value of the parasitic frequency deviation of the instrument signal generator GK4 high-21A in either CW - less than 1 ∙ 10 -4 ƒ;
Output from the uncalibrated output at the end of the high-frequency cable in the continuous mode, - at least 5 ∙ 10 -3 W;
The instrument signal generator GK4 high-21A provides a calibrated output power in the range of 1 ∙ 10 -5 W-1 ∙ 10 -14 watts. In this case:
- Error of the reference level calibrated output power of 1 ∙ 10 -5 W at matched load (VSWR less than 1.4) on the connector OUT. GSS BX. ANALYSIS. SPECTRUM AND wavemeters in the continuous generation of the operating temperature range is less than ± 1,5 dB;
- Error of easing on a scale calibrated attenuator at the site 0dB-70dB in the operating temperature range does not exceed δA1 in dB, defined by the formula:
δA1 = ± (0,8 +0,003 A)
where A - the division of the scale in dB attenuator;
Attenuator setting error signal generator unit GK4 high-21A in the area of 70 dB-90 dB is not standardized;
- The additional error of the output power at low value does not exceed the value of δ A2 in dB, defined by the formula:
δA2 = ± 1 ∙ 10 (A-70) / 20,
where A - the division of the scale in dB attenuator;
Changing the reference level output device when the ambient temperature every 10 ° C - no more than ± 0,6 dB;
Instability of the reference power level in normal conditions after 30 minutes for any subsequent samoprogreva 15 minutes - not exceed ± 0,1 dB;
The extra-warm-up high-frequency generator GK4-21A after the adjustment by frequency - no more than 5 minutes;
Changing the reference level output voltage changes 220V 50Hz to ± 22V, 127V at 50Hz ± 12,7 V, 220 to 400 Hz ± 11V and 115V 400 Hz at ± 5,75 V - no more than ± 0,05 dB;
Bandwidth reduction level of the second and third harmonic relative to the fundamental frequency of the signal level - no less than 30dB;
Changing the level of an output signal at the transition from continuous to pulsed mode - does not exceed ± 0,5 dB;
In PWM mode, the internal launch of a high-frequency signal generator unit GK4-21A produces high-frequency pulses with the following parameters:
- Pulse duration of 0.5 ms, 4 ms, set the length of the error - no more than ± 25%;
- The duration of the rise and fall of high-frequency pulse output - less than 0.2 ms and 0.3 ms, respectively, uneven tops - no more than 25%;
- Pulse repetition frequency of 0.2 kHz, 7 kHz repetition rate setting error - no more than ± 15%;
- Pulses "meander" with a frequency of 1000 Hz ± 15% and duty cycle of 20% ± 2 Hz;
With internal pulse modulation with internal launch of a high-frequency signal generator unit GK4-21A provides timing pulses of either polarity amplitude of at least 50 V to a load of 100 ohms, a fixed duration in the range of 0.1 ms-2mks, a repetition rate of 0.2 kHz, 7 kHz;
External release of a high-frequency signal generator unit GK4-21A pulse is positive or negative duration of 0.2 ms-4mks, a repetition rate of 0.2 kHz, 7 kHz, amplitude 20V-100V;
Parameters of output pulses at high frequency external trigger is the same as for domestic start-up;
The delay in the output of high-frequency pulses of the external trigger is within 1.5 ms-150mks;
Error in setting the delay time - no more than ± 25%;
Note. Setting value is the delay should not exceed half of the pulse;
The instrument signal generator GK4 high-21A is provided by an external modulation:
- Pulses of positive or negative polarity 1mks-20mks duration and frequency of 0.2 kHz, 7 kHz, 30V-70V amplitude, rise time <0.2 ms, the duration of the recession ≤ 0,25 ms, uneven tops no more than 25%;
The rise and decay of high-frequency pulse output - less than 0.2 ms and 0.3 ms, respectively, uneven tops no more than 25%;
- Bursts and frequency of 0.2 kHz, 7 kHz, consisting of 2 ... 5 pulses, positive or negative, the amplitude of 30V-70V, 4-lasting 1mks, 5mks. The smallest interval between pulses 2mks;
Device GK4 high frequency signal generator 21A-measuring external high-frequency power within 1mW-200mW, wherein:
- Accuracy of power 1mW-6mVt not exceed ± 1,5 dB;
- Accuracy of the attenuation of a fixed attenuator with the high number 1 cable and a separate cable does not exceed ± 0,5 dB;
Input VSWR power meter instrument signal generator GK4 high-21A - not more than 1.8;
50 ohm VSWR calibrated instrument output cable is connected - no more than 1.8;
The instrument signal generator GK4 high-21A provides a measure of the frequency integrated wavemeter, with:
- Frequency measurement error of continuous signals does not exceed ± 0,05% at work in the operating temperature range, taking into account the temperature correction;
- Accuracy up to 30MHz frequency difference does not exceed ± 0,5 MHz;
Allowed to increase the frequency difference between the measurement error is not more than a 2-point range to ± 0,75 MHz;
Wavemeter sensitivity by measuring the frequency of the external signal of continuous high-frequency vibrations up to 200 mW at a deviation indicator arrow on half of its scale ≤ 20 mW;
Field strength produced by a high-frequency signal generator unit GK4-21A in the surrounding area, with an output 1 ∙ 10 -14 W at 1m distance from the unit is not more than 1 ∙ 10 -4 V / m;
The instruments provide an analysis of the spectrum of high-frequency pulses with a duration of 0.5 ms-4mks and frequency of 0.4 kHz, 7 kHz, average power of 1 mW, 200 mW, in which:
- The sensitivity of the spectrum analyzer ≤ 1 mW average power;
- Error in determining the width of the main lobe of the spectrum of high-frequency pulse does not exceed ± 15% with pulse duration of 0.5 ms-2, 5mks;
- Error in determining the balance of the principal and the first side lobe of the spectrum of high-frequency pulses of less than ± 15% (at the level of the lateral lobes of at least 0.25 on the level of the ground);
The instrument signal generator GK4 high-21A provides a measurement bandwidth of the receiver in the range of 0.75 MHz to 4 MHz;
The error of measurement bandwidth - no more than ± 20%;
Note. The output from the receiver or filter must be at least 0.5;
The instrument signal generator GK4 high-21A provides its specifications within the established norms of the time after samoprogreva for 30 minutes;
The device retains its specifications within the established norms in their power from AC 220V ± 22V, 50 Hz ± 0,1 Hz or 115 V ± 5,75 V 220V ± 11V frequency of 400 Hz ± 12 Hz and harmonic content up to 5%;
The power consumed by the device signal generator GK4 high-21A from the network at rated voltage - not exceeding 400 V ∙ A;
The instrument high-frequency generator GK4-21A allows continuous operation for 8 hours a day while maintaining their technical specifications within the established norms;
Mean time between failures - not less than 600 hours;
Overall dimensions:
- The measuring unit RIP-IT - no more 445x260x436mm;
- Each block of the high-7M RIP, RIP-10M-15M SSD - no more 445h260h436mm;
Weight:
- The measuring unit RIP-IT - no more than 28kg;
- Each block of the high-7M RIP, RIP-10M-15M SSD - no more than 23kg.

Photos: GK4-21A

GK4-21A generator image.
GK4-21A generator image.
GK4-21A front view.
GK4-21A front view.
GK4-21A side view.
GK4-21A side view.
GK4-21A rear view.
GK4-21A rear view.
GK4-21A side view.
GK4-21A side view.
GK4-21A overhead view.
GK4-21A overhead view.
GK4-21A bottom view.
GK4-21A bottom view.


Яндекс цитирования