Radiometric devices >> sets of measures

E9-85

7 000 RUB
2 200 UAH

Online prices include VAT.

For order is possible effectively delivery of any quantity.

Sets exemplary waveguide loads E9-85

Also, this device can be called: E985, E9 85, E 9-85, E 9 85.

E9-85 sets exemplary waveguide loads are designed for broadband matching output waveguide devices.

Scope:

- To set up, calibration and testing of automatic gauges VSWR;

- For verification of measuring lines and other gauges VSWR;

- To measure the directivity of directional couplers, waveguide to control passthrough devices of various types.

The kit consists of exemplary waveguide load: E9-85/1, E9-85/2, E9-85/3, E9-85/4, E9-85/5.

Specifications E9-85:

Frequency range - from 12,05 GHz to 17,44 GHz.

Cross section of rectangular waveguides - 16×8 mm.

Main technical characteristics:

Rated value

VSWR

Change

VSWR

Deviation

from medium

VSWR values

Error checking

loads based on the reflection of flange connections

2 2% 2% 1%
1,4 1,5% 1,5% 1%
1,14 1,25% 1,25% 0,5%
1,05 1% 1% 0,5%

VSWR absorbing wedges - no more than 1,015.

Uptime - not less than 30000 cycles.

Constancy reflection coefficient module load is within the error checking.

Settings are retained after loads of climatic and mechanical tests.

The manufacturer provides a guarantee of compliance sets loads E9-85 all the technical requirements provided the customer terms and conditions of use, storage and transportation, installed documentation manuals.