Laboratory equipment >> dosimeters, radiometers

DKS-04

19 999 RUB
8 399 UAH
available:
4 pieces

Online prices include VAT.

For order is possible effectively delivery of any quantity.
    Dosimeter DKS-04
    Also, this device can be called: DKS 04, DKS04, DKSO-4, DKSO 4, DKSO4.
    
   
DKS-04 dosimeter is intended for the detection and measurement of thermal neutron flux density, X-ray, gamma and beta radiation, as well as to measure the exposure dose and exposure dose of X-ray and gamma radiation.
    Dosimeters DKS-04 have a portable design.
    
    Specifications dosimeters:
 
    Energy flux of thermal neutrons light, hard beta radiation, X-ray or gamma radiation to trigger audible and visual alarm - over 80 fDzh (0.5 MeV).
    Range of photon energies to measure the exposure dose and exposure dose of X-rays and gamma rays - from 8 to 483 fDzh fDzh (from 0.05 MeV to 3.0 MeV).
    Changing the range of 8 fDzh to 483 fDzh (from 0.05 MeV to 3.0 MeV) - not more than 25%.
    
    Measuring range of the exposure dose (MED) of X-ray and gamma radiation - from 7.16×10-12 A/kg to 7.16×10-8 A/kg (0.1 mR/h to 999.9 mR/h).
    Error of measurement of the exposure dose - ±(19+10/P)%, where P - the exposure dose.
    Measuring range of exposure dose (ED) X-ray and gamma-radiation dose rate from 7.16 in×10-12 A/kg to 7.16×10-8 A/kg (0.1 mR/h to 999.9 mR/h) - from 2.58×10-7 C/kg to 1.032×10-3 C/kg (from 1 mR to 4096 mR).

    Step between the thresholds - 2.58×10-7 C/kg (1 mR).
    Alarm on exceeding each of the 4096's rapids.
    Maximum permissible relative error of measurement of ED - no more than 20%.

    Sound and light alarm at short (less than 2 min) exposure of the exposure dose to 7.16×10-7 A/kg (100 R/h).
    Anisotropy dosimeter DKS-04 with the angle of incidence of the radiation flux of from 0° to 180° relative to the main axis of the detector - not more than ±50% relative to the values ​​measured at an angle of 90°.
    The level of audio signal supplied to the dosimeter - not less than 95 dB.

    Food - from independent power supply voltage of 8 V to 9.5 V.
    Power consumed in the absence of the dosimeter of ionizing radiation - less than 10 mW.
    Warm-up time - no more than 1 min.

    Setting time dosimeter measurement of the exposure dose of X-ray and gamma radiation - less than 5 seconds.
    Time of continuous operation DKS-04:
    - Not less than 12 hours, while the instability of the readings is not greater than ±5%;
    - Not less than 100 hours in the absence of ionizing radiation and background radiation level outside of not more than 3.58×10-12 A/kg (50 mR/hr).
    Maximum permissible additional error dosimeter when the supply voltage to 0.5 V and -1.0 V from the nominal value of 9 - no more than ±10%.

    Dimensions dosimeters DKS-04:
    - Dosimeter - 36x66x150 mm;
    - Body dosimeter 28x64x148 mm;
    - Dosimeter package 70x200x220 mm.

    DKS-04 Weight:
    - Dosimeter - not more than 0.25 kg;
    - Dosimeter package - no more than 0.65 kg.

    MTBF - not less than 2500 hours.
    Specified service life - at least 8 years.
 
    Operating conditions:
 
    Ambient temperature - +10° C to +35° C.
    Relative humidity at +25° C - up to 80%.
    Atmospheric pressure - from 84 kPa to 106.7 kPa (630 mm Hg. Art. 800 mm Hg. Art.)
    
   
Dosimeter DKS-04 is used for individual radiation monitoring laboratories of research institutes and enterprises in various jobs related to ionizing radiation.

Photos: DKS-04

DKS-04 device image.
DKS-04 device image.
DKS-04 accessories and spare parts.
DKS-04 accessories and spare parts.
DKS-04 front view.
DKS-04 front view.
DKS-04 side view.
DKS-04 side view.
DKS-04 rear view.
DKS-04 rear view.
DKS-04 side view.
DKS-04 side view.
DKS-04 overhead view.
DKS-04 overhead view.
DKS-04 bottom view.
DKS-04 bottom view.


Яндекс цитирования