Laboratory equipment >> tensile testing machines

2055 R-0,5 Breaking car 2055 P-0, 5.

2055 R-0,5

For order is possible effectively delivery of any quantity.
Breaking car 2055 R-0,5 (2055 R0,5, 2055 R 0,5)
 
Breaking car 2055 R-0,5, is designed for tensile testing of samples of plastics, ferrous and nonferrous metals at normal temperatures.
 
Technical characteristics of the devices bursting machines 2055 R-0,5:
 
Maximum limit load - 5kN;
Measuring ranges for the speed of the active capture step off the drive X10:
- 1 range - 0mm/min-10mm/min;
- 2 range - 10mm/min-100mm/min;
Measuring ranges for the speed of the active capture device breaking car 2055 R-0,5 step-enabled drive X10:
- 1 range - 10mm/min-100mm/min;
- 2 range - 100mm/min-1000mm/min;
Scale measuring the velocity of the active capture device breaking machine in 2055 R-0,5:
- A range of - 0.1 mm / min;
- 2 range - 1.0 mm / min;
- 3 range - 10.0 mm / min;
- 4 range - 100.0 mm / min;
The limits of acceptable relative error in measuring the speed of movement when working during the active seizure - ± 5%;
No-load speed of the active capture (reverse) unit tensile machine 2055 R-0,5 - at least 1000ob/min;
Measuring ranges and record the load must be:
- 1 range - 20N-100N;
- 2 range - 50N-250N;
- 3 range - 100N-500N;
- 4 range - 200N-1000H;
- 5 Range - 500N-2500N;
- 6 Range - 1000H-5000N;
The number of bits of digital reading device meter unit breaking load washer 2055 R-0,5 - 4;
The nominal unit price of the least significant digit load meter reading device must be:
- A range of - 0.1 N;
- 2 range - 0,025 H;
- 3 range - 0.05 h;
- 4 range - 0.1 N;
- 5 Range - 0.25 h;
- 6 Range - 0.5 N;
The limits of acceptable relative error siloizmeritelya at work during the (loading) of the device breaking car 2055 R-0,5, must be - ± 1%;
The scale readings in the range of measurement machines - should not exceed 1% of the measured load;
The difference in readings between the forward and reverse strokes in the range of measurement - should not exceed 2% of the measured load;
The number of bits of digital reading device measuring displacement of the active capture device breaking machine 2055 R-0,5 - 5;
The nominal unit price of the least significant digit meter reading device moving the active capture - 0.1 mm;
The limits of acceptable absolute error of the meter reading device moving the active capture device breaking car 2055 F-0, 5 - ± 0,5 mm;
The limits of acceptable relative error diagrammatic apparatus for recording load - ± 2% of reading;
The scale of the instrument record breaking car 2055 R-0,5, move the active capture - 50:1, 25:1, 10:1, 5:1, 2:1, 1:1, 1:2.5, 1:1.5 ;
The limits of acceptable error diagrammatic apparatus for recording movement of the active capture, expressed in an appropriate scale, must not exceed:
- At a scale of up to 50:1 recording - up to 30 mm ± 1 mm;
- At the scale of recording up to 100:1 - up to 30 mm ± 2 mm;
- The measured value - more than 30% ± 3%;
Height of space under a tensile test, including the capture of the active stroke - 1000mm;
The width of working space between racks appliance breaking car 2055 R-0,5 - 400 mm;
Power consumption - 0.8 kW;
Dimensions of the machine - 850h645h2080mm;
Dimensions of control panel - 600h735h2200mm;
Weight of the device breaking car 2055 R-0,5 - 330kg;
Weight control unit - 280kg;
It is possible to conduct tests on the machine for compression, bending and low-cycle fatigue. The machine is designed for use in laboratory-type facilities;
The manufacturer guarantees the quality of the unit tensile machine 2055 R-0,5 technical requirements specifications, subject to the consumer the terms and conditions of storage, transportation, installation, operation, established specifications and operational documentation.